Optimizing the operation of a high resolution vertical Johann spectrometer using a high energy fluorescer x-ray source
Authors:
- 1 October 2010
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 81 (10) , 10E537
Abstract
This paper describes the operation and testing for a vertical Johann spectrometer (VJS) operating in the 13 keV range. The spectrometer is designed to use thin curved mica crystals or thick germanium crystals. The VJS must have a resolution of or better to measure the Doppler broadening of highly ionized krypton and operate at a small x-ray angle in order to be used as a diagnostic in a laser plasma target chamber. The VJS was aligned, tested, and optimized using a fluorescer type high energy x-ray (HEX) source located at National Security Technologies (NSTec), LLC, in Livermore, CA. The HEX uses a 160 kV x-ray tube to excite fluorescence from various targets. Both rubidium and bismuth fluorescers were used for this effort. This presentation describes the NSTec HEX system and the methods used to optimize and characterize the VJS performance.
Keywords
This publication has 6 references indexed in Scilit:
- Measuring Curved Crystal Performance for a High-Resolution, Imaging X-Ray SpectrometerX-Ray Optics and Instrumentation, 2010
- Vertical dispersion Johann x-ray spectrometer with asymmetrically cut crystalReview of Scientific Instruments, 2004
- Spectroscopic Determination of Dynamic Plasma Gradients in Implosion CoresPhysical Review Letters, 2002
- High-luminosity, high-resolution, x-ray spectroscopy of laser-produced plasma by vertical-geometry Johann spectrometerReview of Scientific Instruments, 1997
- Diagnosis of High-Temperature Implosions Using Low- and High-Opacity Krypton LinesJournal of X-Ray Science and Technology, 1996
- A modified scheme of Johann spectrographReview of Scientific Instruments, 1995