A new method of measuring in vivo sound speed in the reflection mode
Authors:
- 1 February 1988
- journal article
- research article
- Published by Wiley in Journal of Clinical Ultrasound
- Vol. 16 (2) , 87-93
Abstract
This paper describes a new method for in vivo sound measurement in the reflection mode. The mean sound speed between the reflector and linear array transducer is measured using the following three parameters: time of flight, time of flight difference, and distance between two receiver elements. To detect time of flight, the system delay‐line time compensator is adjusted to obtain the sharpest reflector image.This method was evaluated in vivo in human livers, specifically 26 normal, 27 cirrhotic, and 15 fatty livers. The mean sound speed between diaphragm and the transducer was obtained. The measured sound speed was significantly higher in cirrhotic livers and lower in fatty livers.Keywords
This publication has 6 references indexed in Scilit:
- The butterfly sign: an indicator of fluid within both the greater peritoneal cavity and the lesser omental bursa.Journal of Ultrasound in Medicine, 1983
- Measurement of velocity of propagation from ultrasonic pulse-echo dataUltrasound in Medicine & Biology, 1982
- Acoustic properties of normal and cancerous human liver—I. Dependence on pathological conditionUltrasound in Medicine & Biology, 1981
- Comprehensive compilation of empirical ultrasonic properties of mammalian tissuesThe Journal of the Acoustical Society of America, 1978
- Algebraic Reconstruction of Spatial Distributions of Acoustic Velocities in Tissue from Their Time-of-Flight Profiles1975
- Tables of the Speed of Sound in WaterThe Journal of the Acoustical Society of America, 1959