Technical note: A comparison of in‐house 3D‐printed and commercially available patient‐specific skin collimators for use with electron beam therapy
Open Access
- 26 April 2024
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 25 (5) , e14366
Abstract
Purpose: Skin collimation is a useful tool in electron beam therapy (EBT) to decrease the penumbra at the field edge and minimize dose to nearby superficial organs at risk (OARs), but manually fabricating these collimation devices in the clinic to conform to the patient's anatomy can be a difficult and time intensive process. This work compares two types of patient‐specific skin collimation (in‐house 3D printed and vendor‐provided machined brass) using clinically relevant metrics.Methods: Attenuation measurements were performed to determine the thickness of each material needed to adequately shield both 6 and 9 MeV electron beams. Relative and absolute dose planes at various depths were measured using radiochromic film to compare the surface dose, flatness, and penumbra of the different skin collimation materials.Results: Clinically acceptable thicknesses of each material were determined for both 6 and 9 MeV electron beams. Field width, flatness, and penumbra results between the two systems were very similar and significantly improved compared to measurements performed with no surface collimation.Conclusion: Both skin collimation methods investigated in this work generate sharp penumbras at the field edge and can minimize dose to superficial OARs compared to treatment fields with no surface collimation. The benefits of skin collimation are greatest for lower energy electron beams, and the benefits decrease as the measurement depth increases. Using bolus with skin collimation is recommended to avoid surface dose enhancement seen with collimators placed on the skin surface. Ultimately, the appropriate choice of material will depend on the desire to create these devices in‐house or outsource the fabrication to a vendor.Keywords
This publication has 15 references indexed in Scilit:
- A quick guide on implementing and quality assuring 3D printing in radiation oncologyJournal of Applied Clinical Medical Physics, 2023
- Three-Dimensionally Printed On-Skin Radiation Shields Using High-Density FilamentPractical Radiation Oncology, 2020
- Three‐dimensional printing in radiation oncology: A systematic review of the literatureJournal of Applied Clinical Medical Physics, 2020
- Additive manufacturing in radiation oncology: a review of clinical practice, emerging trends and research opportunitiesInternational Journal of Extreme Manufacturing, 2020
- Material matters: Analysis of density uncertainty in 3D printing and its consequences for radiation oncologyMedical Physics, 2018
- Potential of 3D printing technologies for fabrication of electron bolus and proton compensatorsJournal of Applied Clinical Medical Physics, 2015
- Evaluation of a commercially‐available block for spatially fractionated radiation therapyJournal of Applied Clinical Medical Physics, 2010
- AAPM code of practice for radiotherapy accelerators: Report of AAPM Radiation Therapy Task Group No. 45Medical Physics, 1994
- Clinical electron‐beam dosimetry: Report of AAPM Radiation Therapy Committee Task Group No. 25Medical Physics, 1991
- Lead shielding thickness for dose reduction of 7- to 28-MeV electronsMedical Physics, 1975