Comparison of Standard- and Low-Radiation-Dose CT for Quantification of Emphysema

Authors:

Abstract
OBJECTIVE. This study was performed to compare standard- and low-radiation-dose techniques in the CT quantification of emphysema.MATERIALS AND METHODS. The study population consisted of 36 men and 20 women who were current or former heavy smokers and underwent standard-dose (effective tube current, 100-250 mAs) chest CT at our institution within 6 months of having undergone low-dose (effective tube current, 30-60 mAs) chest CT. All CT scans were reconstructed at 5-mm slice thickness with a smooth filter. CT-measured lung volume, mean and median lung attenuation, and percentage of lung volume with attenuation lower than multiple thresholds (emphysema index values) were compared by Pearson correlation, two-tailed and paired Student's t tests, and regression analysis.RESULTS. There were no significant differences in mean attenuation (-848 vs -846 H, p > 0.35) for the low dose and the standard dose or in median lung attenuation (-879 vs -878 H, p > 0.66). Low- and standard-dose emphysema indexes were correlat...