Feasibility of a depth camera-based center of mass feedback system for closed-loop functional electrical stimulation therapy and visual feedback balance training
Open Access
- 23 July 2026
- journal article
- Published by IOP Publishing in Biomedical Physics & Engineering Express
- Vol. 12 (4) , 045038
Abstract
Individuals with incomplete spinal cord injury are at an increased risk of falls due to impaired balance control. Our group previously developed a closed-loop visual feedback balance training (VFBT) system combined with functional electrical stimulation (FES) therapy to provide standing balance therapy (FES + VFBT system). However, its clinical use has been limited by the reliance on force plates, which are costly and not widely available. The purpose of this study was to develop and evaluate a clinically accessible FES + VFBT system that uses a depth camera in place of a force plate to provide center of mass (COM) feedback. Ten neurologically intact participants (6 males, 4 female) completed three sets of four balance exercises. The COM estimates from the depth camera-based system showed very good correlation with the force plate-based system ( r AP > 0.85; r ML > 0.7) and low root mean squared error, RMSE, (RMSE AP < 0.009 m; RMSE ML < 0.017 m). The stimulation profiles derived from the COM estimates from the depth camera-based system also closely matched those from the force plate system (RMSE < 1.8 mA). These results demonstrate that the proposed depth camera-based FES + VFBT system performs comparably to the original system, offering a more accessible solution for balance therapy in clinical settings.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada (DG RGPIN-2017- 06790)
- CIHR/NSERC/SSHRC Collaborative Health Research Project
This publication has 29 references indexed in Scilit:
- Development of a closed-loop controller for functional electrical stimulation therapy plus visual feedback balance training for standing balance trainingMedical Engineering & Physics, 2024
- Concurrent validity of evaluating knee kinematics using Kinect system during rehabilitation exerciseMedicine in Novel Technology and Devices, 2021
- Comparison of body sway measured by a markerless low-cost motion sensor and by a force plateResearch on Biomedical Engineering, 2021
- Control of standing balance at leaning postures with functional neuromuscular stimulation following spinal cord injuryMedical & Biological Engineering & Computing, 2017
- Upper-limb stroke rehabilitation using electrode-array based functional electrical stimulation with sensing and control innovationsMedical Engineering & Physics, 2016
- Posturography Platform and Balance Control Training and Research System Based on FES and Muscle Synergies2015
- Accuracy of the Microsoft Kinect sensor for measuring movement in people with Parkinson's diseaseGait & Posture, 2014
- Influence of visual inputs on quasi-static standing postural steadiness in individuals with spinal cord injuryGait & Posture, 2013
- Biomechanics and Motor Control of Human Movement2009
- Role of electrical stimulation for rehabilitation and regeneration after spinal cord injury: an overviewEuropean Spine Journal, 2008