DensiProbe Spine
Authors:
- Ronald SchwynRSRonald Schwyn
- Bas Van der PolBPBas Van der Pol
- Markus WindolfMWMarkus Windolf
- Paul F. HeiniPHPaul F. Heini
- Lorin M. BennekerLBLorin M. Benneker
- 1 March 2010
- journal article
- Published by Wolters Kluwer Health in Spine
- Vol. 35 (6) , 607-612
Abstract
Cadaver study. To determine bone strength in vertebrae by measuring peak breakaway torque or indentation force using custom-made pedicle probes. Screw performance in dorsal spinal instrumentation is dependent on bone quality of the vertebral body. To date no intraoperative measuring device to validate bone strength is available. Destructive testing may predict bone strength in transpedicular instrumentations in osteoporotic vertebrae. Insertional torque measurements showed varying results. Ten human cadaveric vertebrae were evaluated for bone mineral density (BMD) measurements by quantitative computed tomography. Peak torque and indentation force of custom-made probes as a measure for mechanical bone strength were assessed via a transpedicular approach. The results were correlated to regional BMD and to biomechanical load testing after pedicle screw implementation. Both methods generated a positive correlation to failure load of the respective vertebrae. The correlation of peak breakaway torque to failure load was r = 0.959 (P = 0.003), therewith distinctly higher than the correlation of indentation force to failure load, which was r = 0.690 (P = 0.040). In predicting regional BMD, measurement of peak torque also performed better than that of indentation force (r = 0.897 [P = 0.002] vs. r = 0.777 [P = 0.017]). Transpedicular measurement of peak breakaway torque is technically feasible and predicts reliable local bone strength and implant failure for dorsal spinal instrumentations in this experimental setting.Keywords
This publication has 18 references indexed in Scilit:
- Osteoporosis: strategies for prevention and managementBest Practice & Research Clinical Rheumatology, 2007
- Mechanical torque measurement predicts load to implant cut-out: a biomechanical study investigating DHS® anchorage in femoral headsArchives of orthopaedic and trauma surgery, 2006
- Epidemiology, treatment and costs of osteoporosis in Germany—the BoneEVA StudyOsteoporosis International, 2006
- The bone quality framework: Determinants of bone strength and their interrelationships, and implications for osteoporosis managementClinical Therapeutics, 2005
- The significance of radiolucent zones surrounding pedicle screwsThe Journal of Bone and Joint Surgery. British volume, 2004
- A prospective study of discordance in diagnosis of osteoporosis using spine and proximal femur bone densitometryOsteoporosis International, 2003
- The relationship between the structural and orthogonal compressive properties of trabecular boneJournal of Biomechanics, 1994
- Complications Associated with the Technique of Pedicle Screw Fixation A Selected Survey of ABS MembersSpine, 1993
- Transpedicular screw-rod fixation of the lumbar spine: operative technique and outcome in 104 casesJournal of Neurosurgery, 1992
- An Experimental Study on Transpedicular Screw Fixation in Relation to Osteoporosis of the Lumbar SpineSpine, 1991