A Case Series of 35 Hip Revisions for Adverse Reactions to Metal Debris following Cormet Hip Resurfacing
- 31 July 2014
- journal article
- research article
- Published by SAGE Publications in HIP International
- Vol. 24 (6) , 575-581
Abstract
The primary aim of this study was to analyse our painful metal-on-metal (MoM) hip resurfacing revisions with evidence of ‘adverse reactions to metal debris’ (ARMD). In our series of 35 revisions the median whole blood Cobalt levels were 58 nmols/l (range 12-1407 nmols/l), and whole blood Chromium levels were 73 nmols/l (range 2-353 nmols/l). Thirty-four of our 35 patients had abnormal imaging on Ultrasound scanning (USS). The mean histological Campbell grading of ARMD was 4, and ranged from 0-9. The mean Oxford Hip Score (OHS) increased from 19 pre-revision (range 4-46) to 33 post-revision surgery (range 23-47). We found no correlation between the preoperative metal ion levels, and the severity of the disease or the outcome. Pain following hip resurfacing may arise from a number of causes and when groin pain arises in conjunction with abnormal cross sectional imaging we have offered our patients revision surgery regardless of raised metal ions or grossly abnormal imaging, with good results.Keywords
This publication has 19 references indexed in Scilit:
- European multidisciplinary consensus statement on the use and monitoring of metal-on-metal bearings for total hip replacement and hip resurfacingOrthopaedics & Traumatology: Surgery & Research, 2013
- Adverse reactions to metal debris: histopathological features of periprosthetic soft tissue reactions seen in association with failed metal on metal hip arthroplastiesJournal of Clinical Pathology, 2012
- Adverse reaction to metal debris following hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2011
- The relationship between head-neck ratio and pseudotumour formation in metal-on-metal resurfacing arthroplasty of the hipThe Journal of Bone and Joint Surgery. British volume, 2010
- Birmingham hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2010
- Optimal acetabular orientation for hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2010
- Histological Features of Pseudotumor-like Tissues From Metal-on-Metal HipsClinical Orthopaedics and Related Research, 2010
- Analysis of wear of retrieved metal-on-metal hip resurfacing implants revised due to pseudotumoursThe Journal of Bone and Joint Surgery. British volume, 2010
- Risk factors for inflammatory pseudotumour formation following hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2009
- Pseudotumours associated with metal-on-metal hip resurfacingsThe Journal of Bone and Joint Surgery. British volume, 2008