Pulse-Shape Discrimination of 3-in Size NaI:Tl,⁶Li Scintillator Coupled With SiPM Array

Authors:

Abstract
Single crystals of NaI:Tl, 6 Li, a newly developed neutron-gamma scintillator, have satisfactory n/γ discrimination capability and ease of scaling up. Using a SiPM instead of a PMT as a readout for detector miniaturization has attracted considerable attention. Herein, we systematically compare the gamma spectroscopy and n/γ discrimination performance of a 3-inch NaI:Tl, 6 Li single crystal using R6233-100 PMT and two types of SiPM arrays as readouts. Using the PMT as a readout, a satisfactory energy resolution of 8.6% at 662 keV and an excellent figure of merit (FoM) of 4.36 at 1 GHz sampling rate can be achieved. Using the two types of SiPM arrays as readouts, the detectors have good linearity with energy resolutions of 14.4% and 10.6% at 662 keV, and FoM values greater than 3 at 1 GHz sampling rate. Notably, at low sampling rates, the detector based on a 3-inch NaI:Tl, 6 Li single crystal and SiPM array can maintain a very beneficial pulse-shape discrimination (PSD) performance, such as a FoM value of up to 2.73 at a sampling rate of 20 MHz.