A flow cytometer chip fabricated in 65-nm standard CMOS technology embedded into a microfluidic channel is presented. The chip detects cells specifically tagged with magnetic labels by demodulating the change in coupling between an excitation coil and a pickup coil of a differential spiral transformer. Operation of the flow cytometer cartridge is verified for the detection of SKBR-3 breast cancer cells tagged with manganese ferrite magnetic labels using EpCAM-biotin-streptavidin chemistry. Magnetic labels of different materials are classified by the difference in the phase of their complex susceptibility.