Sensing Aberration by the Automatic Implantable Cardioverter Defibrillator During Intraoperative Testing
Authors:
- 1 March 1988
- journal article
- case report
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 11 (3) , 331-335
Abstract
A 50‐year‐old man underwent replacement of his automatic implantable cardioverter defibrillator (AICD) because a magnet test revealed severe battery depletion. He had had his unit implanted 18 months previously after an episode of sudden cardiac death. He had documented torsades de pointes and inducible ventricular tachycardia, confirmed by electrophysiologic study. Before a new unit was implanted thresholds were measured by an external cardioverter defibrillator. Ventricular fibrillation (VF) was induced by alternating current through a standard, line‐operated battery charger with stimulation delivered to the epicardium via rate‐sensing electrodes. VF was allowed to continue for 10 seconds before shock was delivered. Termination o/VF required 15 joules, which was higher than that required at initial implantation 18 months earlier. The new pulse generator was activated/or testing and VF was again induced. The AICD discharged after 12.3 seconds. Prior to wound closing, the AICD was deactivated by magnet. Instead of R‐wave synchronous beeping tones during deactivation, double beeping tones were heard. Electrogram recordings revealed abnormalities of the T‐wave and ST segment of the rate‐sensing electrodes, which were the cause of the tone irregularities. Stabilization of the T‐wave and ST segment occurred within 8 minutes and the tones became normal. The procedure was then completed.Keywords
This publication has 5 references indexed in Scilit:
- Failure of the automatic implantable defibrillator to detect ventricular fibrillationThe American Journal of Cardiology, 1986
- The Automatic Implantable Cardioverter-Defibrillator: Efficacy, Complications, and Device FailuresAnnals of Internal Medicine, 1986
- The automatic implantable cardioverter-defibrillator: An overviewJACC, 1985
- Practical aspects of automatic cardioverter/defibrillator implantationAmerican Heart Journal, 1984
- The automatic implantable defibrillator: Local ventricular bipolar sensing to detect ventricular tachycardia and fibrillationThe American Journal of Cardiology, 1983