Multicenter Evaluation Study Comparing a New Factory-Calibrated Real-Time Continuous Glucose Monitoring System to Existing Flash Glucose Monitoring System
Authors:
- 11 August 2021
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 17 (1) , 208-213
Abstract
This study reports a clinical evaluation of AiDEX CGM system featuring a 14-day sensor, real-time glucose monitoring and factory-calibration. A multicenter, prospective, masked clinical study was conducted at with a total of 120 participants. Each participant wore 4 studied sensors and had one in-clinic visit for venous blood reference tests. 40 out of the 120 participants wore additional Abbott Libre sensors and performed at least 7 capillary BG tests daily for additional reference and comparison. Continuous glucose error grid analysis (CG-EGA) showed that AiDEX and Abbott Libre had good agreement with venous blood glucose, with 98.69% and 98.96% accurate readings, respectively. Overall MARD of AiDEX CGM systems was 9.08% when compared to venous blood reference and 10.1% when compared to finger capillary BG reference.Keywords
Funding Information
- Microtech Medical (Hangzhou) Co. Ltd.
This publication has 9 references indexed in Scilit:
- Accuracy of a Factory-Calibrated, Real-Time Continuous Glucose Monitoring System During 10 Days of Use in Youth and Adults with DiabetesDiabetes Technology & Therapeutics, 2018
- Allergic contact dermatitis caused by isobornyl acrylate in Freestyle® Libre, a newly introduced glucose sensorContact Dermatitis, 2017
- Cutaneous adverse events related to FreeStyle Libre deviceThe Lancet, 2017
- Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trialThe Lancet, 2016
- The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring SystemDiabetes Technology & Therapeutics, 2015
- Continuous Glucose Monitoring in the Subcutaneous Tissue over a 14-Day Sensor Wear PeriodJournal of Diabetes Science and Technology, 2013
- Evaluating the Accuracy of Continuous Glucose-Monitoring SensorsDiabetes Care, 2004
- A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.Diabetes Care, 2000
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987