Mural destabilization after aneurysm treatment with a flow-diverting device: a report of two cases
Authors:
- 1 March 2011
- journal article
- case report
- Published by BMJ in Journal of NeuroInterventional Surgery
- Vol. 3 (2) , 167-171
Abstract
Background Flow-diverting stents have demonstrated great promise for the treatment of cerebral aneurysms; however, clinical experience with the devices remains very preliminary. We present two cases of spontaneous delayed complications—one fatal aneurysm rupture and one symptomatic increase in aneurysm volume—following the treatment of intradural aneurysms with the Pipeline Embolization Device (PED). Presentation/intervention Two patients with unruptured, intradural aneurysms of the carotid artery underwent uneventful treatment with the PED (eV3, Irvine, California, USA). One patient, with a giant aneurysm of the carotid terminus, experienced worsening headache 5 days after the procedure and ultimately collapsed and became unresponsive. CT of the head demonstrated acute subarachnoid and intraventricular hemorrhage. The patient died the following day. A second patient with a large left posterior communicating artery aneurysm presented with progressive memory loss 3 months after PED reconstruction of the carotid artery. Although serial CT angiograms showed progressive thrombosis of the aneurysm to near-complete occlusion, MR of the brain demonstrated marked interval growth of the collective aneurysm–intra-aneurysmal thrombus mass with extensive edema throughout the adjacent left temporal lobe. Conclusions Flow-diverting devices have demonstrated tremendous promise for the treatment of complex, unruptured cerebral aneurysms. However, experience with this novel approach to aneurysm treatment is preliminary and the consequences of its application within the cerebrovasculature remain incompletely defined. Mural destabilization resulting in delayed, spontaneous, aneurysm growth and/or rupture may occur in the days to weeks following the application of flow-diverting devices to treat previously unruptured intracranial aneurysms. A better understanding of the incidence and etiology of these complications is essential for this technology to be optimally applied.Keywords
This publication has 17 references indexed in Scilit:
- Intra-Aneurysmal Thrombosis as a Possible Cause of Delayed Aneurysm Rupture after Flow-Diversion TreatmentAmerican Journal of Neuroradiology, 2010
- Early fatal hemorrhage after endovascular cerebral aneurysm treatment with a flow diverter (SILK-Stent)Neuroradiology, 2010
- The mechanical role of thrombus on the growth rate of an abdominal aortic aneurysmJournal of Vascular Surgery, 2010
- CURATIVE ENDOVASCULAR RECONSTRUCTION OF CEREBRAL ANEURYSMS WITH THE PIPELINE EMBOLIZATION DEVICENeurosurgery, 2009
- CURATIVE RECONSTRUCTION OF A GIANT MIDBASILAR TRUNK ANEURYSM WITH THE PIPELINE EMBOLIZATION DEVICENeurosurgery, 2009
- DEFINITIVE RECONSTRUCTION OF CIRCUMFERENTIAL, FUSIFORM INTRACRANIAL ANEURYSMS WITH THE PIPELINE EMBOLIZATION DEVICENeurosurgery, 2008
- The ELANA technique: Constructing a high flow bypass using a non-occlusive anastomosis on the ICA and a conventional anastomosis on the SCA in the treatment of a fusiform giant basilar trunk aneurysmActa Neurochirurgica, 2004
- Influence of intraluminal thrombus on structural and cellular composition of abdominal aortic aneurysm wallJournal of Vascular Surgery, 2003
- Growth of Thrombus may be a Better Predictor of Rupture than Diameter in Patients with Abdominal Aortic AneurysmsEuropean Journal of Vascular and Endovascular Surgery, 2000
- Surgery in Spontaneous Subarachnoid HaemorrhageBMJ, 1956