Evaluation of 24-h urine containers for urine copper measurement by inductively coupled plasma mass spectrometry
- 6 April 2020
- journal article
- research article
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 57 (3) , 246-248
Abstract
Objective: To determine the necessity for acid-washed containers for 24-h urine copper analysis. Methods: Copper solutions, with concentrations relevant to clinical decision limits, were prepared by spiking both assay diluent and unknown urine samples with the copper calibrator. Samples were split between plain and acid-washed 24-h urine containers, and copper analysis was performed using inductively coupled mass spectrometry (ICP-MS). Results: Measurement of copper in both spiked diluent and spiked urine samples showed minimal concentration bias between acid-washed and plain 24-h urine containers. Conclusions: Acid-washed containers are not required for the measurement of copper in 24-h urine samples.Keywords
This publication has 1 reference indexed in Scilit:
- Comparison of Sample Preservation Methods for Clinical Trace Element Analysis by Inductively Coupled Plasma Mass SpectrometryAmerican Journal of Clinical Pathology, 2005