Real‐time x‐ray fluoroscopy‐based catheter detection and tracking for cardiac electrophysiology interventions
Authors:
- YingLiang MaYMYingLiang Ma
- Nicolas GoginNGNicolas Gogin
- Pascal CathierPCPascal Cathier
- R. James HousdenR. James Housden
- Geert GijsbersGGGeert Gijsbers
- Michael CooklinMCMichael Cooklin
- Mark O'NEillMOMark O'NEill
- Jaswinder GillJGJaswinder Gill
- C. Aldo RinaldiCRC. Aldo Rinaldi
- Reza RazaviRRReza Razavi
- Kawal S. RhodeKawal S. Rhode
Open Access
- 11 June 2013
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 40 (7) , 071902
Abstract
Purpose:: X‐ray fluoroscopically guided cardiac electrophysiology (EP) procedures are commonly carried out to treat patients with arrhythmias. X‐ray images have poor soft tissue contrast and, for this reason, overlay of a three‐dimensional (3D) roadmap derived from preprocedural volumetric images can be used to add anatomical information. It is useful to know the position of the catheter electrodes relative to the cardiac anatomy, for example, to record ablation therapy locations during atrial fibrillation therapy. Also, the electrode positions of the coronary sinus (CS) catheter or lasso catheter can be used for road map motion correction.Methods:: In this paper, the authors present a novel unified computational framework for image‐based catheter detection and tracking without any user interaction. The proposed framework includes fast blob detection, shape‐constrained searching and model‐based detection. In addition, catheter tracking methods were designed based on the customized catheter models input from the detection method. Three real‐time detection and tracking methods are derived from the computational framework to detect or track the three most common types of catheters in EP procedures: the ablation catheter, the CS catheter, and the lasso catheter. Since the proposed methods use the same blob detection method to extract key information from x‐ray images, the ablation, CS, and lasso catheters can be detected and tracked simultaneously in real‐time.Results:: The catheter detection methods were tested on 105 different clinical fluoroscopy sequences taken from 31 clinical procedures. Two‐dimensional (2D) detection errors of 0.50 ± 0.29, 0.92 ± 0.61, and 0.63 ± 0.45 mm as well as success rates of 99.4%, 97.2%, and 88.9% were achieved for the CS catheter, ablation catheter, and lasso catheter, respectively. With the tracking method, accuracies were increased to 0.45 ± 0.28, 0.64 ± 0.37, and 0.53 ± 0.38 mm and success rates increased to 100%, 99.2%, and 96.5% for the CS, ablation, and lasso catheters, respectively. Subjective clinical evaluation by three experienced electrophysiologists showed that the detection and tracking results were clinically acceptable.Conclusions:: The proposed detection and tracking methods are automatic and can detect and track CS, ablation, and lasso catheters simultaneously and in real‐time. The accuracy of the proposed methods is sub‐mm and the methods are robust toward low‐dose x‐ray fluoroscopic images, which are mainly used during EP procedures to maintain low radiation dose.Keywords
Funding Information
- EPSRC-GB (WT 088641/Z/09/Z)
This publication has 23 references indexed in Scilit:
- Registration of 3D trans-esophageal echocardiography to X-ray fluoroscopy using image-based probe trackingMedical Image Analysis, 2012
- Modeling and Registration for Electrophysiology Procedures Based on Three-Dimensional ImagingCurrent Cardiovascular Imaging Reports, 2011
- Hybrid echo and x-ray image guidance for cardiac catheterization procedures by using a robotic arm: a feasibility studyPhysics in Medicine & Biology, 2010
- Three-dimensional left atrial and esophagus reconstruction using cardiac C-arm computed tomography with image integration into fluoroscopic views for ablation of atrial fibrillation: Accuracy of a novel modality in comparison with multislice computed tomographyHeart Rhythm, 2008
- Biplane three-dimensional augmented fluoroscopy as single navigation tool for ablation of atrial fibrillation: Accuracy and clinical valueHeart Rhythm, 2008
- Computed tomography-fluoroscopy overlay evaluation during catheter ablation of left atrial arrhythmiaEP Europace, 2008
- Time-resolved three-dimensional imaging of the left atrium and pulmonary veins in the interventional suite—A comparison between multisweep gated rotational three-dimensional reconstructed fluoroscopy and multislice computed tomographyHeart Rhythm, 2008
- Computed Tomography‐Fluoroscopy Image Integration‐Guided Catheter Ablation of Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2007
- Three-dimensional rotational angiography of the left atrium and esophagus—A virtual computed tomography scan in the electrophysiology lab?Heart Rhythm, 2007
- A system for real-time XMR guided cardiovascular interventionIEEE Transactions on Medical Imaging, 2005