Determination of Hg in water and wastewater samples by CV-AAS following on-line preconcentration with silver trap
- 29 August 2010
- journal article
- research article
- Published by Pleiades Publishing Ltd in Journal of Analytical Chemistry
- Vol. 65 (9) , 935-939
Abstract
An online mercury preconcentration and determination system consisting of cold vapor atomic absorption spectrometry (CV-AAS) coupled to a flow injection (FI) method was studied. The method was developed involving the determination of ng/L levels of mercury retained on the silver wool solid sorbent. Experimental conditions such as sample volume, flow rate, stability of the column and effect of foreign ions on the determination of trace amounts of mercury were optimized. The detection limit is 3 ng/L and dynamic range 10–250 ng/L require only 50 mL of sample. The relative standard deviations (RSD) of the determinations are below 4%. The presence of common metal ions, such as K+, Na+, Cu2+, Pb2+, Fe3+, Ni2+, and Mn2+, does not interfere with the measurement of mercury by this method. The method was successfully applied to the determination of mercury in water and wastewater samples.Keywords
This publication has 40 references indexed in Scilit:
- Preparation of the diphenylcarbazone-functionalized silica gel and its application to on-line selective solid-phase extraction and determination of mercury by flow-injection spectrophotometryJournal of Hazardous Materials, 2008
- Determination of methylmercury and inorganic mercury in water samples by slurry sampling cold vapor atomic absorption spectrometry in a flow injection system after preconcentration on silica C18 modifiedTalanta, 2007
- Highly selective determination of trace quantities of mercury in water samples after preconcentration by the cloud-point extraction methodInternational Journal of Environmental Analytical Chemistry, 2006
- Determination of methylmercury and estimation of total mercury in seafood using high performance liquid chromatography (HPLC) and inductively coupled plasma-mass spectrometry (ICP-MS): Method development and validationAnalytica Chimica Acta, 2006
- Alumina modified by dimethyl sulfoxide as a new selective solid phase extractor for separation and preconcentration of inorganic mercury(II)Talanta, 2006
- Sensitized extraction spectrophotometric determination of Hg(II) with dithizone after its flotation as ion-associate using iodide and ferroinTalanta, 2005
- Silica gel functionalized with resacetophenone: synthesis of a new chelating matrix and its application as metal ion collector for their flame atomic absorption spectrometric determinationAnalytica Chimica Acta, 2002
- Determination of Organomercury and Mercury in Environmental Samples by Flow Injection Atomic Fluorescence SpectrophotometryInternational Journal of Environmental Analytical Chemistry, 1996
- Microcolumn Field Sampling and Flow Injection-Atomic Fluorescence Spectrometry for Speciation of Mercury in Surface WatersInternational Journal of Environmental Analytical Chemistry, 1994
- Automated determination of mercury at ultra trace level in waters by gold amalgam preconcentration and cold vapour atomic fluorescence spectrometryAnalytica Chimica Acta, 1993