A Small Surrogate for the Golden Angle in Time-Resolved Radial MRI Based on Generalized Fibonacci Sequences
Authors:
- 18 December 2014
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 34 (6) , 1262-1269
Abstract
In golden angle radial magnetic resonance imaging a constant azimuthal radial profile spacing of 111.246... ° guarantees a nearly uniform azimuthal profile distribution in k-space for an arbitrary number of radial profiles. Even though this profile order is advantageous for various real-time imaging methods, in combination with balanced steady-state free precession (SSFP) sequences the large azimuthal angle increment may lead to strong image artifacts, due to the varying eddy currents introduced by the rapidly switching gradient scheme. Based on a generalized Fibonacci sequence, a new sequence of smaller irrational angles is introduced ( 49.750... ° , 32.039... ° , 27.198... ° , 23.628... ° , ... ). The subsequent profile orders guarantee the same sampling efficiency as the golden angle if at least a minimum number of radial profiles is used for reconstruction. The suggested angular increments are applied for dynamic imaging of the heart and the temporomandibular joint. It is shown that for balanced SSFP sequences, trajectories using the smaller golden angle surrogates strongly reduce the image artifacts, while the free retrospective choice of the reconstruction window width is maintained.Keywords
This publication has 26 references indexed in Scilit:
- Motion corrected compressed sensing for free‐breathing dynamic cardiac MRIMagnetic Resonance in Medicine, 2012
- The influence of radial undersampling schemes on compressed sensing reconstruction in breast MRIMagnetic Resonance in Medicine, 2011
- Temporomandibular DisordersNew England Journal of Medicine, 2008
- Compressed Sensing MRIIEEE Signal Processing Magazine, 2008
- Analysis and compensation of eddy currents in balanced SSFPMagnetic Resonance in Medicine, 2005
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- k-Space weighted image contrast (KWIC) for contrast manipulation in projection reconstruction MRIMagnetic Resonance in Medicine, 2000
- Reduced aliasing artifacts using variable-densityk-space sampling trajectoriesMagnetic Resonance in Medicine, 2000
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Nonlinear total variation based noise removal algorithmsPhysica D: Nonlinear Phenomena, 1992