Ion-Selective Electrode Determination of Sodium and Potassium in Blood and Urine
Authors:
- 1 January 1986
- journal article
- research article
- Published by Taylor & Francis in Analytical Letters
- Vol. 19 (17-18) , 1809-1830
Abstract
The Beckman Electrolyte 2 Analyzer (E2A) has been studied for its response to sodium and potassium content in various laboratory prepared pH-adjusted mixtures of metal chlorides in the absence and presence of albumin and α-, β-and γ -globulins. These mixtures also included MITS, a Mock Ionic Test Serm, containing sodium chloride (15OmM), potassium chloride (4mM), calcium chloride (1mM) and magnesium chloride (1mM). Results for sodium and potassium are often low for these non-clinical standards. Responses have also been studied for patient blood serum and urine samples. Some comparisons have been made with data obtained by flame photometry. The blood serum data obtained with the Electrolyte 2 Analyzer are statistically compared with those obtained by flame photometry.Keywords
This publication has 5 references indexed in Scilit:
- A Comparison of the Measurement of Sodium and Potassium by Flame Photometry and Ion-Selective ElectrodeAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1985
- Sampling and storage of blood for determination of ionized calciumScandinavian Journal of Clinical and Laboratory Investigation, 1985
- Practical Suggestions for Testing Automatic Equipment Based on Ion-Selective Electrodes1984
- Neutral Carrier Based Ion-Selective Electrodes1983
- Transient phenomena at glass electrodesAnalytical Chemistry, 1967