Intraoperative zero-heat-flux thermometry overestimates nasopharyngeal temperature by 0.39 °C: an observational study in patients undergoing congenital heart surgery
Open Access
- 10 August 2024
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 39 (1) , 205-215
Abstract
During surgery for congenital heart disease (CHD) temperature management is crucial. Vesical (Tves) and nasopharyngeal (TNPH) temperature are usually measured. Whereas Tves slowly responds to temperature changes, TNPH carries the risk of bleeding. The zero-heat-flux (ZHF) temperature monitoring systems SpotOn™ (TSpotOn), and Tcore™ (Tcore) measure temperature non-invasively. We evaluated accuracy and precision of the non-invasive devices, and of Tves compared to TNPH for estimating temperature. In this prospective observational study in pediatric and adult patients accuracy and precision of TSpotOn, Tcore, and Tves were analyzed using the Bland-Altman method. Proportion of differences (PoD) and Lin´s concordance correlation coefficient (LCC) were calculated. Data of 47 patients resulted in sets of matched measurements: 1073 for TSpotOn vs. TNPH, 874 for Tcore vs. TNPH, and 1102 for Tves vs. TNPH. Accuracy was − 0.39 °C for TSpotOn, -0.09 °C for Tcore, and 0.07 °C for Tves. Precisison was between − 1.12 and 0.35 °C for TSpotOn, -0.88 to 0.71 °C for Tcore, and − 1.90 to 2.05 °C for Tves. PoD ≤ 0.5 °C were 71% for TSpotOn, 71% for Tcore, and 60% for Tves. LCC was 0.9455 for TSpotOn, 0.9510 for Tcore, and 0.9322 for Tves. Temperatures below 25.2 °C (TSpotOn) or 27.1 (Tcore) could not be recorded non-invasively, but only with Tves. Trial registration German Clinical Trials Register, DRKS00010720.Keywords
Funding Information
- Herzzentrum Göttingen
This publication has 46 references indexed in Scilit:
- Non-invasive zero-heat-flux technology compared with traditional core temperature measurements in the emergency departmentThe American Journal of Emergency Medicine, 2020
- Intraoperative zero-heat-flux thermometry overestimates esophageal temperature by 0.26 °C: an observational study in 100 infants and young childrenJournal of Clinical Monitoring and Computing, 2020
- Accuracy and precision of zero-heat-flux temperature measurements with the 3M™ Bair Hugger™ Temperature Monitoring System: a systematic review and meta-analysisJournal of Clinical Monitoring and Computing, 2020
- Accuracy of the non-invasive Tcore™ temperature monitoring system to measure body core temperature in abdominal surgeryJournal of Clinical Monitoring and Computing, 2019
- Zero-heat-flux core temperature monitoring system: an observational secondary analysis to evaluate agreement with naso-/oropharyngeal probe during anesthesiaJournal of Clinical Monitoring and Computing, 2019
- The focus of temperature monitoring with zero-heat-flux technology (3M Bair-Hugger): a clinical study with patients undergoing craniotomyJournal of Clinical Monitoring and Computing, 2018
- Evaluation of the novel non-invasive zero-heat-flux Tcore™ thermometer in cardiac surgical patientsJournal of Clinical Monitoring and Computing, 2018
- The accuracy of a disposable noninvasive core thermometerCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2013
- Nasal Temperature can be used as a Reliable Surrogate Measure of Core TemperatureJournal of Clinical Monitoring and Computing, 2008
- Perioperative Maintenance of Normothermia Reduces the Incidence of Morbid Cardiac EventsJAMA, 1997