Local impedance‐guided radiofrequency ablation with standard and high power: Results of a preclinical investigation
- 5 July 2021
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 32 (8) , 2060-2068
Abstract
Background: Local impedance (LI) drop measured with microfidelity electrodes embedded in the tip of an ablation catheter accurately reflects tissue heating during radiofrequency (RF) ablation. Previous studies found 15–30 Ω LI drops created successful lesions, while more than 40 Ω drops were associated with steam pops. The objective of this study was to evaluate the safety and efficacy of LI‐guided ablation using standard (30 W) and high‐power (50 W) in a preclinical model.Methods: RF lesions were created in explanted swine hearts (n = 6) to assess the feasibility of LI‐guided ablation by targeting 10, 20, or 30 Ω (n = 20/group) drops. Subsequently, LI‐guided ablation was evaluated in a chronic animal model (n = 8 Canines, 25–29 kg, 30/50 W). During the index procedure point‐by‐point intercaval line ablation and left inferior pulmonary vein (PV) isolation were performed. RF duration was at the operators’ discretion but discontinued early if a 15–30 Ω drop was achieved. Operators attempted to avoid LI drops of more than 40 Ω. At 1‐month, durable conduction block was evaluated with electroanatomic mapping followed by necropsy and histopathology.Results: In explanted tissue, terminating ablation at 10, 20, or 30 Ω LI drops created statistically larger lesions (p < .05; 1.8 [1.6–2.4] mm, 3.3 [3.0–3.7] mm; 4.9 [4.3–5.5] mm). LI‐guided high‐power ablation in vivo significantly reduced RF duration per application compared to standard‐power (p < .05; intercaval: 8.9 ± 5.2 vs. 18.1 ± 11.0 s, PV: 9.6 ± 5.4 vs. 23.2 ± 10.3 s). LI drops of 15–40 Ω were more readily achievable for high‐power (90.1%, 318/353) than standard‐power (71.7%, 243/339). All intercaval lines and PV isolations were durable (16/16) at 1‐month. Necropsy revealed no major collateral injury to the pericardium, phrenic nerve, esophagus, or lungs. There was no pericardial effusion, stroke, tamponade, or PV stenosis. Vagal nerve injury was found in two 30 W animals after using 19.7 ± 13.9 and 19.5 ± 11.8 s RF applications.Conclusion: LI‐guided ablation was found to be safe and efficacious in a chronic animal model. High‐power ablation more readily achieved more than 15 Ω drops, reduced RF duration compared with standard‐power, and had no major RF collateral injury.Keywords
Funding Information
- Boston Scientific Corporation
This publication has 26 references indexed in Scilit:
- Novel method for electrode-tissue contact measurement with multi-electrode cathetersEP Europace, 2017
- Factors Associated With Periesophageal Vagal Nerve Injury After Pulmonary Vein Antrum IsolationJournal of the American Heart Association, 2014
- Initial impedance decrease as an indicator of good catheter contact: Insights from radiofrequency ablation with force sensing cathetersHeart Rhythm, 2014
- Electrical Reconnection After Pulmonary Vein Isolation Is Contingent on Contact Force During Initial TreatmentCirculation: Arrhythmia and Electrophysiology, 2013
- Use of electrical coupling information in AF catheter ablation: A prospective randomized pilot studyHeart Rhythm, 2013
- Area Under the Real‐Time Contact Force Curve (Force–Time Integral) Predicts Radiofrequency Lesion Size in an In Vitro Contractile ModelJournal of Cardiovascular Electrophysiology, 2010
- Updated Worldwide Survey on the Methods, Efficacy, and Safety of Catheter Ablation for Human Atrial FibrillationCirculation: Arrhythmia and Electrophysiology, 2010
- First in Human Validation of Impedance‐Based Catheter Tip‐to‐Tissue Contact Assessment in the Left AtriumJournal of Cardiovascular Electrophysiology, 2009
- Aging Dilates Atrium and Pulmonary VeinsChest, 2008
- Anatomic Relations Between the Esophagus and Left Atrium and Relevance for Ablation of Atrial FibrillationCirculation, 2005