Establishment of Outcome-Related Analytic Performance Goals
- 1 May 2010
- journal article
- review article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 56 (5) , 714-722
Abstract
Background: Accrediting organizations require laboratories to establish analytic performance criteria that ensure their tests provide results of the high quality required for patient care. However, the procedures for instituting performance criteria that are directly linked to the needs of medical practice are not well established, and therefore alternative strategies often are used to create and implement surrogate performance standards.Content: We reviewed 6 approaches for establishing outcome-related analytic performance goals: (a) limits defined by regulations and external assessment programs, (b) limits based on biologic variation, (c) limits based on surveys of clinicians about their needs, (d) limits based on effects on guideline driven medical decisions, (e) limits based on analysis of patterns for ordering follow-up clinical tests, and (f) limits based on formal medical decision models. Performance criteria were tabulated for 12 common chemistry analytes and 4 routine hematology tests.Conclusions: There is no consensus currently about the preferred methods for establishing medically necessary analytic performance limits. The various methods reviewed give considerably different performance limits. The analytic performance limits claimed by a laboratory should correspond to those limits that can be reliably maintained based on validated QC monitoring systems. These limits generally are larger than the observed CVs and bias parameters collected for assay validation. There is a major need for increased communication among laboratorians and clinicians on this topic, especially when the analytic performance limits that can be consistently maintained by a laboratory are inconsistent with the expectations of health care providers.Keywords
This publication has 32 references indexed in Scilit:
- Clinical decision-making as the basis for assessing agreement between measures of the International Normalized RatioJournal of Thrombosis and Haemostasis, 2009
- CDC project on standardizing steroid hormone measurementsSteroids, 2008
- Extensions to decision curve analysis, a novel method for evaluating diagnostic tests, prediction models and molecular markersBMC Medical Informatics and Decision Making, 2008
- CommentaryClinical Chemistry, 2008
- Integration of Meta-analysis and Economic Decision Modeling for Evaluating Diagnostic TestsMedical Decision Making, 2008
- What Information Should Manufacturers Provide on Their Procedures?Clinical Chemistry, 2006
- Introduction: Strategies to set global quality specifications in laboratory medicineScandinavian Journal of Clinical and Laboratory Investigation, 1999
- Analytic bias specifications based on the analysis of effects on performance of medical guidelinesScandinavian Journal of Clinical and Laboratory Investigation, 1999
- Clinical assessment of haemoglobin values by general practitioners related to analytical and biological variationScandinavian Journal of Clinical and Laboratory Investigation, 1991
- Analytical goals for the acceptance of common reference intervals for laboratories throughout a geographical areaScandinavian Journal of Clinical and Laboratory Investigation, 1988