Is Ablation of Atrial Flutter Always Safe?

Authors:

Abstract
Background: Radiofrequency ablation of typical atrial flutter is largely used and is considered as safe. The purpose of the study was to evaluate the prevalence and the causes of severe adverse event (AE) following atrial flutter ablation. Methods: Ablation of typical flutter was performed by conventional method with an 8‐mm‐tip electrode catheter, a maximum power of 70 W, and a maximum target temperature of 70° for 60 seconds in 883 patients, (685 males and 198 females aged from 18 to 93 years [64 ± 11.5]; 664 had heart disease [HD]). Results: AE occurred in 44 patients (5%). AE was life threatening in 14 patients: poorly tolerated bradycardia (transient complete atrioventricular block [AVB] or sinus bradycardia [SB] Conclusions: AE following atrial flutter ablation occurred in 5% of patients. Most of them are avoidable by control of anticoagulants and arrest of rate‐control drugs used to slow the rate of atrial flutter. (PACE 2012; 35:1061–1066)