Usefulness of Rate Responsive Atrial Pacing in Patients with Sick Sinus Syndrome
Authors:
- Rinya KatoRKRinya Kato
- Hiroshi HayashiHHHiroshi Hayashi
- Shoji NodaSNShoji Noda
- Iwao SotobataISIwao Sotobata
- Mitsuya MuraseMMMitsuya Murase
- 1 January 1989
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 12 (P1) , 16-24
Abstract
We studied heart rate response and blood lactate changes during treadmill exercise testing in 10 patients with sick sinus syndrome (SSS) and normal atrioventricular (AV) conduction in whom an activity detecting multi-programmable rate responsive pacemaker was implanted for atrial pacing (AAI pacing). Treadmill exercise testing was performed twice in each patient under the two different AAI pacing modes: rate responsive function off (fixed rate 60 pulses per minute [ppm]; RR-off) and rate responsive function on (rate response 10, activity threshold medium, basic rate 60 ppm, and maximum rate 150 ppm; RR-on). Chronotropic response and blood lactate change during exercise were compared among patients under each pacing mode and eight healthy volunteers of ages matching those of the patients. In patients under the RR-off pacing mode (RR-off patients), chronotropic response was lower (P less than 0.01) and blood lactate level was higher (P less than 0.05) than those in patients under the RR-on pacing mode (RR-on patients) and volunteers during exercise. In RR-on patients, however, both chronotropic response and lactate level changes during exercise were comparatively improved, with no significant differences between RR-on patients and volunteers. These data suggested that improvement of chronotropic response by rate responsive pacing, in comparison with nonrate responsive pacing, increased the blood supply to the working muscles and resulted in less lactate production during exercise. It was concluded that rate responsive AAI pacing which could maintain AV synchrony and normal intraventricular conduction was an optimal physiological pacing mode for SSS patients with normal AV conduction.Keywords
This publication has 17 references indexed in Scilit:
- Measurement of Right Ventricular Blood Temperature During Exercise as a Means of Rate Control in Physiological Pacemakers.Pacing and Clinical Electrophysiology, 1986
- Atrial versus ventricular pacing in sinus node disease: A treatment comparison studyAmerican Heart Journal, 1986
- Rate‐Responsive Pacing: Clinical ExperiencePacing and Clinical Electrophysiology, 1985
- Current concepts in sick sinus syndrome. II. ECG manifestation and diagnostic and therapeutic approachesArchives of internal medicine (1960), 1985
- Activity‐Sensing, Rate‐Responsive Pacing: Improvement in Myocardial Performance with ExercisePacing and Clinical Electrophysiology, 1985
- Increasing Cardiac Rate by Tracking the Respiratory RatePacing and Clinical Electrophysiology, 1984
- A Theoretical Right Atrial Pressure Feedback Heart Rate Control System to Restore Physiologic Control to the Rate‐limited HeartPacing and Clinical Electrophysiology, 1984
- Stability of AV conduction in sick sinus node syndrome patients with implanted atrial pacemakersAmerican Heart Journal, 1984
- Results, Problems and Perspectives with the Autoregulating PacemakerPacing and Clinical Electrophysiology, 1983
- A method of permanent transvenous implantation of an atrial electrodeAmerican Heart Journal, 1973