Performance of intermittently scanned continuous glucose monitoring systems in people with type 1 diabetes: A pooled analysis
Open Access
- 19 December 2021
- journal article
- research article
- Published by Wiley in Diabetes, Obesity and Metabolism
- Vol. 24 (3) , 522-529
Abstract
Aims: To conduct a pooled analysis to assess the performance of intermittently scanned continuous glucose monitoring (isCGM) in association with the rate of change in sensor glucose in a cohort of children, adolescents, and adults with type 1 diabetes. Material and Methods: In this pooled analysis, isCGM system accuracy was assessed depending on the rate of change in sensor glucose. Clinical studies that have been investigating isCGM accuracy against blood glucose, accompanied with collection time points were included in this analysis. isCGM performance was assessed by means of median absolute relative difference (MedARD), Parkes error grid (PEG) and Bland‐Altman plot analyses. Results: Twelve studies comprising 311 participants were included, with a total of 15 837 paired measurements. The overall MedARD (interquartile range) was 12.7% (5.9‐23.5) and MedARD differed significantly based on the rate of change in glucose ( P < 0.001). An absolute difference of −22 mg/dL (−1.2 mmol/L) (95% limits of agreement [LoA] 60 mg/dL (3.3 mmol/L), −103 mg/dL (−5.7 mmol/L)) was found when glucose was rapidly increasing (isCGM glucose minus reference blood glucose), while a −32 mg/dL (1.8 mmol/L) (95% LoA 116 mg/dL (6.4 mmol/L), −51 mg/dL (−2.8 mmol/L)) absolute difference was observed in periods of rapidly decreasing glucose. Conclusions: The performance of isCGM was good when compared to reference blood glucose measurements. The rate of change in glucose for both increasing and decreasing glucose levels diminished isCGM performance, showing lower accuracy during high rates of glucose change.Keywords
This publication has 33 references indexed in Scilit:
- Evaluation of subcutaneous glucose monitoring systems under routine environmental conditions in patients with type 1 diabetesDiabetes, Obesity and Metabolism, 2017
- Accuracy and Longevity of an Implantable Continuous Glucose Sensor in the PRECISE Study: A 180-Day, Prospective, Multicenter, Pivotal TrialDiabetes Care, 2016
- Comparison of Two Continuous Glucose Monitoring Systems, Dexcom G4 Platinum and Medtronic Paradigm Veo Enlite System, at Rest and During ExerciseDiabetes Technology & Therapeutics, 2016
- Accuracy of Continuous Glucose Monitoring (CGM) during Continuous and High-Intensity Interval Exercise in Patients with Type 1 Diabetes MellitusNutrients, 2016
- Head-to-head comparison between flash and continuous glucose monitoring systems in outpatients with type 1 diabetesJournal of Endocrinological Investigation, 2016
- Rate-of-Change Dependence of the Performance of Two CGM Systems During Induced Glucose SwingsJournal of Diabetes Science and Technology, 2015
- Perceived Accuracy in Continuous Glucose MonitoringJournal of Diabetes Science and Technology, 2014
- Technical Aspects of the Parkes Error GridJournal of Diabetes Science and Technology, 2013
- Precision and accuracy of blood glucose measurements using three different instrumentsDiabetic Medicine, 2012
- Continuous glucose monitoring system during physical exercise in adolescents with type 1 diabetesActa Paediatrica, 2011