Studies of the interactions of an MRI system with the shielding in a combined PET/MRI scanner
Authors:
- 11 December 2009
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 55 (1) , 265-280
Abstract
A positron emission tomography (PET) system or 'insert' has been constructed for placement and operation in the bore of a small animal magnetic resonance imaging (MRI) scanner to allow simultaneous MR and PET imaging. The insert contains electronics, components with a variety of magnetic properties and large continuous sheets of metal--all characteristics of an object that should, by conventional wisdom, never be placed in the bore of an MR scanner, especially near the imaging volume. There are a variety of ways the two systems might be expected to interact that could negatively impact the performance of either or both. In this article, the interaction mechanisms, particularly the impact of the PET insert and shielding on MR imaging, are defined and explored. Additionally, some of the difficulties in quantifying errors introduced into the MR images as a result of the presence of the PET components are demonstrated. Several different approaches are used to characterize image artifacts and determine optimal placement of the shielding. Data are also presented that suggest ways the shielding could be modified to reduce errors and enable placement closer to the isocenter of the magnet.Keywords
This publication has 26 references indexed in Scilit:
- Recent developments in PET detector technologyPhysics in Medicine & Biology, 2008
- Simultaneous in vivo positron emission tomography and magnetic resonance imagingProceedings of the National Academy of Sciences of the United States of America, 2008
- Multimodality imaging of structure and functionPhysics in Medicine & Biology, 2008
- Design of field-cycled magnetic resonance systems for small animal imagingPhysics in Medicine & Biology, 2006
- Computer simulation studies of the effects of dynamic shimming on susceptibility artifacts in EPI at high fieldJournal of Magnetic Resonance, 2005
- Characterization and reduction of gradient‐induced eddy currents in the RF shield of a TEM resonatorMagnetic Resonance in Medicine, 2002
- Simple Correction Method fork-Space Trajectory Deviations in MRIJournal of Magnetic Resonance, 1998
- Positron emission tomography within a magnetic field using photomultiplier tubes and lightguidesPhysics in Medicine & Biology, 1995
- Automatic, localized in Vivo adjustment of all first‐and second‐order shim coilsMagnetic Resonance in Medicine, 1993
- Chemical shift imaging in nuclear magnetic resonance: A comparison of methodsConcepts in Magnetic Resonance, 1993