Repeatability of Corneal Thickness Measured Using an Oculus Pentacam
- 1 March 2009
- journal article
- research article
- Published by Wiley in Optometry and Vision Science
- Vol. 86 (3) , 266-272
Abstract
Purpose.: To evaluate the repeatability of corneal thickness measurements over time, obtained with an Oculus Pentacam. Methods.: Measurements of corneal thickness (apical and peripheral) were performed at three different sessions in one eye of 23 healthy subjects, using an Oculus Pentacam instrument. In each session, three consecutive measurements were taken (without realignment) by the same observer. Measurements were performed, immediately after a blink. Results.: Mean values obtained were in agreement with published literature values for central corneal thickness (mean ± SEM: 546.23 ± 5.24 μm for first session over the first 1 min, 545.13 ± 5.49 μm for second session 1 h later, and 543.23 ± 4.96 μm for the third session measured 1 week later). The data showed an increase in corneal thickness towards the periphery with a decrease in intraobserver reproducibility. The confidence limits in the Bland and Altman charts for pachymetry at the corneal apex ranged between −10.91 μm and 13.00 μm within the first session measured over the first 1 min, −13.43 μm to 11.23 μm between the first and second sessions after 1 h and −11.99 μm to 10.02 μm between the first and third sessions after 1 week. The confidence limits were found to increase towards the peripheral cornea. There was no significant difference in within‐subjects variance between the three time scales (p > 0.05 for all parameters measured). Conclusion.: Our data showed that the Oculus Pentacam provides repeatable measures of corneal thickness when measured within a few seconds, after 1 h, and 1 week.Keywords
This publication has 32 references indexed in Scilit:
- Pentacam pachometry: comparison with non‐contact specular microscopy on the central cornea and inter‐session repeatability on the peripheral corneaClinical and Experimental Optometry, 2007
- Diurnal Variation of Central Corneal Thickness and Goldmann Applanation Tonometry Estimates of Intraocular PressureJournal of Glaucoma, 2007
- Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetryJournal of Cataract & Refractive Surgery, 2005
- Central corneal thickness of Caucasians, Chinese, Hispanics, Filipinos, African Americans, and Japanese in a glaucoma clinicOphthalmology, 2004
- Test–Retest Reliability of Clinical Shack-Hartmann MeasurementsInvestigative Ophthalmology & Visual Science, 2004
- The effects of long-term contact lens wear on corneal thickness, curvature, and surface regularityOphthalmology, 2000
- Epithelial and corneal thickness measurements by in vivo confocal microscopy through focusing (CMTF)Current Eye Research, 1997
- Diurnal variations in human corneal thickness.British Journal of Ophthalmology, 1996
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- An analysis of variance test for normality (complete samples)Biometrika, 1965