Correction of Footdrop Due to Multiple Sclerosis Using the STIMuSTEP Implanted Dropped Foot Stimulator
Authors:
- Paul N. TaylorPTPaul N. Taylor
- Ingrid A. Wilkinson HartIHIngrid A. Wilkinson Hart
- Mansoor S. KhanMKMansoor S. Khan
- Diana E.M. Slade-SharmanDSDiana E.M. Slade-Sharman
- 1 September 2016
- journal article
- Published by Consortium of Multiple Sclerosis Centers in International Journal of MS Care
- Vol. 18 (5) , 239-247
Abstract
Background: Footdrop is a significant problem in multiple sclerosis, reducing the safety and efficiency of walking. Functional electrical stimulation (FES) can produce dorsiflexion, correcting footdrop. The purpose of this retrospective analysis of clinical study data was to compare the effect of external and implanted FES devices for the correction of footdrop. Methods: External FES was used for a minimum of 6 months before implantation. Walking performance was assessed using 10-m walking speed, 3-minute walking distance, the Physiological Cost Index, and health- and device-related quality of life and device-use questionnaires. Assessments were made before implantation and a mean (SD) of 128 (24) days after surgery, with additional walking speed measurements at 3 years. Results: Twenty-three people with multiple sclerosis received the STIMuSTEP implant. Both devices enabled statistically significant increases in walking speed and walking distance, with a strong trend toward a reduced Physiological Cost Index, indicating that walking required less effort (P = .07). Both devices improved device-related quality of life. Walking speed gain with FES was maintained at 3 years. Three implants failed after falls, and there was one case of neuropraxia. The implant was used more days per week and was quicker to put on each day than the external FES device. Conclusions: The STIMuSTEP implanted dropped foot stimulator is an effective long-term intervention for the correction of footdrop.Keywords
This publication has 30 references indexed in Scilit:
- The long-term cost-effectiveness of the use of Functional Electrical Stimulation for the correction of dropped foot due to upper motor neuron lesionJournal of Rehabilitation Medicine, 2013
- Poster 358: Comparison of Functional Electric Stimulation Neuroprosthesis and Ankle Foot Orthosis in Persons With Multiple SclerosisPM&R, 2010
- Impact on activities of daily living using a functional electrical stimulation device to improve dropped foot in people with multiple sclerosis, measured by the Canadian Occupational Performance MeasureMultiple Sclerosis Journal, 2010
- Long-Term Therapeutic and Orthotic Effects of a Foot Drop Stimulator on Walking Performance in Progressive and Nonprogressive Neurological DisordersNeurorehabilitation and Neural Repair, 2009
- The effect of functional electrical stimulation on the physiological cost of gait in people with multiple sclerosisMultiple Sclerosis Journal, 2008
- Meaningful Change and Responsiveness in Common Physical Performance Measures in Older AdultsJournal of the American Geriatrics Society, 2006
- The Orthotic Effect of Functional Electrical Stimulation on the Improvement of Walking in Stroke Patients with a Dropped Foot: A Systematic ReviewArtificial Organs, 2004
- The significant change for the Timed 25-Foot Walk in the Multiple Sclerosis Functional CompositeMultiple Sclerosis Journal, 2000
- Validation of the Short Form 36 (SF-36) Health Survey Questionnaire Among Stroke PatientsStroke, 1996
- Classification of Walking Handicap in the Stroke PopulationStroke, 1995