Validation of current density measurements with a B-dot probe
Authors:
- 1 October 2001
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 72 (10) , 4002-4004
Abstract
The time varying magnetic flux density in an electrodeless high frequency discharge was measured with a bare wire B-dot probe immersed in plasma. Based on B-dot measurements, the spatial distribution of the current density was determined and integrated over the cross section of the discharge to determine the total discharge current. The total current determined from B-dot measurements was compared with total current measured externally with a current transformer. Agreement between the two methods was very good over a wide range of gas pressures and discharge currents. This work validates that a small, immersed B-dot probe does not significantly perturb the discharge current and can accurately determine current density in a low-pressure gas discharge.Keywords
This publication has 8 references indexed in Scilit:
- Electric field in inductively coupled gas dischargesJournal of Applied Physics, 2001
- Theoretical and experimental investigations of the electromagnetic field within a planar coil, inductively coupled RF plasma sourcePlasma Sources Science and Technology, 1998
- The electric field and current density in a low-pressure inductive discharge measured with different B-dot probesJournal of Applied Physics, 1997
- Plasma properties determined with induction loop probes in a planar inductively coupled plasma sourceJournal of Applied Physics, 1996
- Magnetic field distribution measurements in a low-pressure inductive dischargeJournal of Applied Physics, 1995
- Measurement of plasma current distribution using magnetic probesMeasurement Science and Technology, 1994
- Electromagnetic fields in a radio-frequency induction plasmaJournal of Vacuum Science & Technology A, 1993
- Magnetic Probe ResolutionReview of Scientific Instruments, 1964