A Portable, High-Speed, Vacuum-Outlet GC Vapor Analyzer Employing Air as Carrier Gas and Surface Acoustic Wave Detection
Authors:
- 5 September 2001
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 73 (19) , 4668-4675
Abstract
Vacuum-outlet GC with atmospheric-pressure air as the carrier gas is implemented at outlet pressures up to 0.8 atm using a low-dead-volume polymer-coated surface acoustic wave (SAW) detector. Increases in the system outlet pressure from 0.1 to 0.8 atm lead to proportional increases in detector sensitivity and significant increases in column efficiency. The latter effect arises from the fact that optimal carrier gas velocities are lower in air than in more conventional carrier gases such as helium or hydrogen due to the smaller binary diffusion coefficients of vapors in air. A 12-m-long, 0.25-mm-i.d. tandem column ensemble consisting of 4.5-m dimethyl polysiloxane and 7.5-m trifluoropropylmethyl polysiloxane operated at an outlet pressure of 0.5 atm provides up to 4 × 104 theoretical plates and a peak capacity of 65 (resolution, 1.5) for a 3-min isothermal analysis. At 30 °C, mixtures of vapors ranging in vapor pressure from 8.6 to 76 Torr are separated in this time frame. The SAW detector cell has an internal volume of <2 μL, which allows the use of higher column outlet pressures with minimal dead time. The sensor response is linear with solute mass over at least 2−3 decades and provides detection limits of 20−50 ng for the vapors tested. The combination of atmospheric-pressure air as carrier gas, modest operating pressures, and SAW sensor detection is well-suited for field instrumentation since it eliminates the need for support gases, permits smaller, low-power pumps to be used, and provides sensitivity to a wide range of vapor analytes.Keywords
This publication has 33 references indexed in Scilit:
- A gas chromatography-quartz crystal microbalance for speciation of sulfur compounds in landfill gas2000
- Temperature Programming for High-Speed GCAnalytical Chemistry, 1999
- High-Speed, Vacuum-Outlet GC Using Atmospheric-Pressure Air as Carrier GasAnalytical Chemistry, 1999
- Application of piezoelectric Ru(III)/cryptand-coated quartz crystal gas chromatographic detector for olefinsAnalytica Chimica Acta, 1999
- Effects of Temperature and Humidity on the Performance of Polymer-Coated Surface Acoustic Wave Vapor Sensor ArraysAnalytical Chemistry, 1996
- High-speed GC analysis of VOCs: sample collection and inlet systems. 1.Environmental Science & Technology, 1994
- Sample introduction in high speed capillary gas chromatography; input band width and detection limitsJournal of High Resolution Chromatography, 1987
- Improvement of speed of separation in packed column gas chromatographyAnalytical Chemistry, 1982
- Computer comparisons of variables in capillary gas chromatographyJournal of High Resolution Chromatography, 1982
- On-line computer controlled multiple ion detection in combined gas chromatography-mass spectrometryAnalytical Chemistry, 1973