Third Prize: The Impact of Fluid Environment Manipulation on Shockwave Lithotripsy Artificial Calculi Fragmentation Rates
Authors:
- 1 March 2011
- journal article
- research article
- Published by SAGE Publications in Journal of Endourology
- Vol. 25 (3) , 397-401
Abstract
Background and Purpose: Studies have suggested that shockwave lithotripsy (SWL) stone fragmentation rates can be affected by characteristics of the fluid media surrounding the stone, although evidence to implicate the impact of urine specific gravity (SG) is limited and inconclusive. Our aim is to further explore the impact fluid media and SGs have on stone fragmentation using a variable focus lithotripter. Materials and Methods: Artificial stones were presoaked for 24 hours in urine and then shocked in various fluid media including artificial urine (SG 1.010 control, 1.020, and 1.07), human pooled urine (HPU), degassed HPU, Pentastarch, 100% and 30% contrast, degassed 30% contrast, 100% ethanol, deionized water (dH2O), degassed dH2O, 5% glucose, Ringer lactate, 0.9% saline, glycerol, whole blood, and lubricating gel. After soaking, SWL using the Modulith SLX-F2 electromagnetic lithotripter was performed. Fragments were dried and sieved using a 4-mm diameter opening grid. Fragments >4 mm were weighed and fragmentation coefficients (FCs) calculated (pre-SWL weight – post-SWL weight)/(pre-SWL weight) × 100. Fifteen stones were shocked for each fluid group. Results: Fluid type, viscosity, and degassing all significantly impacted stone fragmentation. While the solutions' SG, per se, did not appear to affect stone fragmentation, the use of degassed 30% contrast significantly improved stone destruction over the SG 1.010 artificial urine control (95.3% vs 71.4, P < 0.01). Furthermore, degassing improved comminution rates by increasing the number of completely fragmented stones (FC = 100%). Using degassed 30% contrast, 12/15 stones were completely fragmented, compared with only 2/15 in the control group (P = 0.007). Among the whole blood, glycerol, and lubricating gel groups, only 1/15, 0/15, and 1/15 stones reached 100% FC respectively in the narrow focus, possibly because of the detrimental impact of increased viscosity. Conclusions: Different fluid media can significantly affect FC in vitro. Among the various fluids tested, degassed 30% contrast significantly increased the FC and total number of completely fragmented stones.Keywords
This publication has 14 references indexed in Scilit:
- Quantitative measurements of acoustic emissions from cavitation at the surface of a stone in response to a lithotripter shock waveThe Journal of the Acoustical Society of America, 2006
- Effect of Urine Specific Gravity on Effectiveness of Shockwave LithotripsyJournal of Endourology, 2005
- Cavitation Bubble Cluster Activity in the Breakage of Kidney Stones by Lithotripter ShockwavesJournal of Endourology, 2003
- Current Management of Urolithiasis: Progress or Regress?Journal of Endourology, 2002
- The role of stress waves and cavitation in stone comminution in shock wave lithotripsyUltrasound in Medicine & Biology, 2002
- IN VITRO COMPARISON OF SHOCK WAVE LITHOTRIPSY MACHINESJournal of Urology, 2000
- A simple artificial urine for the growth of urinary pathogensLetters in Applied Microbiology, 1997
- Quantitation of shock wave cavitation damage in vitroUltrasound in Medicine & Biology, 1997
- Minimal static excess pressure minimises the effect of extracorporeal shock waves on cells and reduces it on gallstonesUltrasound in Medicine & Biology, 1997
- Biomechanical Effects of Extracorporeal Shock Wave Lithotripsy: Relation of Applied Voltage and Shock Wave Number to Cell Viability and Stone Fragmentation*Journal of Endourology, 1990