Cranial Reconstruction with Computer-Generated Hard-Tissue Replacement Patient-Matched Implants: Indications, Surgical Technique, and Long-Term Follow-Up
Authors:
- Barry L. EppleyBEBarry L. Eppley
- Matthew KilgoMKMatthew Kilgo
- John J. ColemanJCJohn J. Coleman
- 1 March 2002
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Plastic and Reconstructive Surgery
- Vol. 109 (3) , 864-871
Abstract
The aim of this clinical study was to evaluate the effectiveness and safety of using computer-generated alloplastic (hard-tissue replacement) implants for the reconstruction of large defects of the upper craniofacial region. Fourteen patients who had large (> 150 cm2) preexisting defects of the cranium or cranio-orbital region underwent surgical reconstruction. Preoperatively, a three-dimensional computed tomographic scan was obtained from which an anatomic model was fabricated. The defect in the model was then used to create an alloplastic (hard tissue-replacement polymer) implant for reconstruction and surgical placement. At the time of surgery, the implant was secured into position with either metal or resorbable fixation. In cases where the frontal sinus was in proximity to the implant, the frontal sinus was either cranialized and covered with a pericranial flap or obliterated with hydroxyapatite cement. In cases that had been previously irradiated or infected, wide bony debridement and coverage with a vascularized muscle was initially performed, followed by implant reconstruction 6 months later. All implants fit easily into the bone defects, and only four (29 percent) required some minor adjustments to complete the fit. All patients healed uneventfully. With a minimum of 1 year follow-up (average, 3 years) in all cases, excellent contours have been maintained and all patients have remained infection-free. In large cranial defects, custom implants fabricated from porous, hydrophilic hardtissue replacement polymer provide an exacting anatomic fit and a solid stable reconstruction. This method of reconstruction in these defects is rapid and exact, and significantly reduces operative time. Critical attention must be paid, however, to management of the frontal sinus and preexisting bone infection and the quality of the overlying soft-tissue cover. (Plast. Reconstr. Surg. 109: 864, 2002.)Keywords
This publication has 14 references indexed in Scilit:
- Re: Producing a Full-Scale Model From Computed Tomographic Data with the Rapid Prototyping Technique Using the Binder Jet Method: A Comparison with the Laser Lithography Method Using a Dry Skull.Journal of Craniofacial Surgery, 2001
- Treatment of Large Complex Cranial Bone Defects by Using Hydroxyapatite Ceramic ImplantsPlastic and Reconstructive Surgery, 1999
- Stereolithographic biomodelling in cranio-maxillofacial surgery: a prospective trialJournal of Cranio-Maxillofacial Surgery, 1999
- Computer-Generated Patient Models for Reconstruction of Cranial and Facial DeformitiesJournal of Craniofacial Surgery, 1998
- Custom-made cast titanium implants produced with CAD/CAM for the reconstruction of cranium defectsInternational Journal of Oral and Maxillofacial Surgery, 1998
- The value of stereolithographic models for preoperative diagnosis of craniofacial deformities and planning of surgical correctionsInternational Journal of Oral and Maxillofacial Surgery, 1998
- Orbital Reconstruction with Individually Copy-Milled Ceramic ImplantsPlastic and Reconstructive Surgery, 1998
- Indications and limitations of three-dimensional models in cranio-maxillofacial surgeryJournal of Cranio-Maxillofacial Surgery, 1998
- Method for Preparing an Exact-Size Model Using Helical Volume Scan Computed TomographyPlastic and Reconstructive Surgery, 1994
- Computer-Designed Prostheses for Orbitocranial ReconstructionPlastic and Reconstructive Surgery, 1988