Revision of metal-on-metal hip arthroplasty in a tertiary center
Authors:
- Alexander D LiddleALAlexander D Liddle
- Keshtra SatchithanandaKSKeshtra Satchithananda
- Johann HenckelJHJohann Henckel
- Shiraz A SabahSSShiraz A Sabah
- Karuniyan V VipulendranKVKaruniyan V Vipulendran
- Angus LewisALAngus Lewis
- John A SkinnerJSJohn A Skinner
- Adam W M MitchellAMAdam W M Mitchell
- Alister J HartAHAlister J Hart
Open Access
- 18 April 2013
- journal article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 84 (3) , 237-245
Abstract
Operative findings during revision of metal-on-metal hip arthroplasty (MOMHA) vary widely and can involve massive soft tissue and bone disruption. As a result, planning of theater time and resources is difficult, surgery is challenging, and outcomes are often poor. We describe our experience with revision of MOMHA and provide recommendations for management. We present the findings and outcomes of 39 consecutive MOMHAs (in 35 patients) revised in a tertiary unit (median follow-up time 30 (12-54) months). The patients underwent a preoperative work-up including CT, metal artifact reduction sequence (MARS) MRI, and blood metal ion levels. We determined 5 categories of failure. 8 of 39 hips had conventional failure mechanisms including infection and impingement. Of the other 31 hips, 14 showed synovitis without significant disruption of soft tissue; 6 had a cystic pseudotumor with significant soft tissue disruption; 7 had significant osteolysis; and 4 had a solid pseudotumor. Each category of failure had specific surgical hazards that could be addressed preoperatively. There were 2 reoperations and 1 patient (2 hips) died of an unrelated cause. Median Oxford hip score (OHS) was 37 (9-48); median change (ΔOHS) was 17 (-10 to 41) points. ΔOHS was similar in all groups-except those patients with solid pseudotumors and those revised to metal-on-metal bearings, who fared worse. Planning in revision MOMHA is aided by knowledge of the different categories of failure to enable choice of appropriate personnel, theater time, and equipment. With this knowledge, satisfactory outcomes can be achieved in revision of metal-on-metal hip arthroplasty.Keywords
This publication has 17 references indexed in Scilit:
- Consensus Statement “Current Evidence on the Management of Metal-on-Metal Bearings”: April 16, 2012HIP International, 2013
- Accelerating failure rate of the ASR total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2011
- Revisions of Metal-on-Metal Hip Resurfacing: Lessons Learned and Improved OutcomeOrthopedic Clinics of North America, 2011
- Magnetic Resonance Imaging Findings in Painful Metal-On-Metal HipsThe Journal of Arthroplasty, 2011
- Failed Metal-on-Metal Hip Arthroplasties: A Spectrum of Clinical Presentations and Operative FindingsClinical Orthopaedics and Related Research, 2010
- Hip resurfacings revised for inflammatory pseudotumour have a poor outcomeThe Journal of Bone and Joint Surgery. British volume, 2009
- Dose-dependent cytotoxicity of clinically relevant cobalt nanoparticles and ions on macrophagesin vitroBiomedical Materials, 2009
- Can the Acetabular Position be Derived from a Pelvic Frame of Reference?Clinical Orthopaedics and Related Research, 2008
- Metal-on-metal hip resurfacing arthroplasty: A review of periprosthetic biological reactionsActa Orthopaedica, 2008
- The use of the Oxford hip and knee scoresThe Journal of Bone and Joint Surgery. British volume, 2007