Analysis of partial volume effects on arterial input functions using gradient echo: A simulation study
Open Access
- 13 April 2009
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 61 (6) , 1300-1309
Abstract
Absolute blood flow and blood volume measurements using perfusion weighted MRI require an accurately measured arterial input function (AIF). Because of limited spatial resolution of MR images, AIF voxels cannot be placed completely within a feeding artery. We present a two‐compartment model of an AIF voxel including the relaxation properties of blood and tissue. Artery orientations parallel and perpendicular to the main magnetic field were investigated and AIF voxels were modeled to either include or be situated close to a large artery. The impact of partial volume effects on quantitative perfusion metrics was investigated for the gradient echo pulse sequence at 1.5 T and 3.0 T. It is shown that the tissue contribution broadens and introduces fluctuations in the AIF. Furthermore, partial volume effects bias perfusion metrics in a nonlinear fashion, compromising quantitative perfusion estimates and profoundly effecting local AIF selection. Magn Reson Med, 2009.Keywords
Funding Information
- Danish National Research Foundation
- the EU IST Program, and The Danish Research Council NABIIT Program
This publication has 34 references indexed in Scilit:
- Contrast agent concentration measurements affecting quantification of bolus‐tracking perfusion MRIMagnetic Resonance in Medicine, 2007
- Partial volume effects on arterial input functions: Shape and amplitude distortions and their correctionJournal of Magnetic Resonance Imaging, 2005
- Arterial input functions for dynamic susceptibility contrast MRI: Requirements and signal optionsJournal of Magnetic Resonance Imaging, 2005
- The impact of partial‐volume effects in dynamic susceptibility contrast magnetic resonance perfusion imagingJournal of Magnetic Resonance Imaging, 2005
- A comparison of T2*-weighted magnitude and phase imaging for measuring the arterial input function in the rat aorta following intravenous injection of gadolinium contrast agentMagnetic Resonance Imaging, 2005
- Defining a local arterial input function for perfusion MRI using independent component analysisMagnetic Resonance in Medicine, 2004
- Influence of Arterial Input Function on Hypoperfusion Volumes Measured With Perfusion-Weighted ImagingStroke, 2004
- Measuring the arterial input function with gradient echo sequencesMagnetic Resonance in Medicine, 2003
- Quantification of Perfusion Using Bolus Tracking Magnetic Resonance Imaging in StrokeStroke, 2002
- Delay and dispersion effects in dynamic susceptibility contrast MRI: Simulations using singular value decompositionMagnetic Resonance in Medicine, 2000