A Colloid Osmometer for Small Fluid Samples
Authors:
- 1 February 1974
- journal article
- Published by Wiley in Acta Physiologica Scandinavica
- Vol. 90 (2) , 485-490
Abstract
A membrane colloid osmometer for measurement on samples down to 5 μl has been developed. Pellucid acryl plastic is used for most parts of the osmometer, allowing visual inspection of the reference fluid chamber. The osmometer base is attached directly to a Hewlett‐Packard pressure transducer, replacing the original dome of the transducer, thereby avoiding extra connections and leakage. The exposed membrane area has a diameter of 3 mm, and the membrane may be changed without completely disassembling the osmometer. With Amicon UM‐10 membranes and 0.9 % saline as reference fluid, the equilibration time for measurements on human sera was 2–4 min. Reduction of equilibration time to about 1 min was obtained by “priming” and flushing the membrane with serum ultrafiltrate. Reproducible measurements were obtained on samples down to 4–5 μl.Keywords
This publication has 7 references indexed in Scilit:
- Protein Concentration of Interstitial Fluid Collected from Rat Skin by a Wick MethodActa Physiologica Scandinavica, 1973
- An improved osmometer for measurement of plasma colloid osmotic pressureMicrovascular Research, 1972
- Measurement of blood plasma colloid osmotic pressure: II. Comparative study of different speciesMicrovascular Research, 1971
- Measurement of blood plasma colloid osmotic pressure: I. Technical aspectsMicrovascular Research, 1971
- Direct continuous recording of plasma colloid osmotic pressure of whole blood.Journal of Applied Physiology, 1968
- Measurement of the Colloid Osmotic Pressure in PatientsScandinavian Journal of Clinical and Laboratory Investigation, 1967
- A Self‐Recording Electronic Osmometer for Quick, Direct Measurement of Colloid Osmotic Pressure in Small Samples1Acta Physiologica Scandinavica, 1961