Signal averaging improves signal-to-noise in OCT images: But which approach works best, and when?
Open Access
- 17 October 2019
- journal article
- research article
- Published by Optica Publishing Group in Biomedical Optics Express
- Vol. 10 (11) , 5755-5775
Abstract
The high acquisition speed of state-of-the-art optical coherence tomography (OCT) enables massive signal-to-noise ratio (SNR) improvements by signal averaging. Here, we investigate the performance of two commonly used approaches for OCT signal averaging. We present the theoretical SNR performance of (a) computing the average of OCT magnitude data and (b) averaging the complex phasors, and substantiate our findings with simulations and experimentally acquired OCT data. We show that the achieved SNR performance strongly depends on both the SNR of the input signals and the number of averaged signals when the signal bias caused by the noise floor is not accounted for. Therefore we also explore the SNR for the two averaging approaches after correcting for the noise bias and, provided that the phases of the phasors are accurately aligned prior to averaging, then find that complex phasor averaging always leads to higher SNR than magnitude averaging.Keywords
Funding Information
- Austrian Science Fund (P25823-B24, P26553-N20)
- H2020 European Research Council (ERC Starting Grant 640396 OPTIMALZ)
This publication has 45 references indexed in Scilit:
- Enhanced Vitreous Imaging in Healthy Eyes Using Swept Source Optical Coherence TomographyPLOS ONE, 2014
- Motion artifact and speckle noise reduction in polarization sensitive optical coherence tomography by retinal trackingBiomedical Optics Express, 2013
- Assessment of Frame-Averaging Algorithms in OCT Image AnalysisOphthalmic Surgery, Lasers and Imaging Retina, 2013
- Efficient reduction of speckle noise in Optical Coherence TomographyOptics Express, 2012
- Birefringence measurement of cornea and anterior segment by office-based polarization-sensitive optical coherence tomographyBiomedical Optics Express, 2011
- Analyses and calculations of noise in optical coherence tomography systemsFrontiers of Optoelectronics in China, 2008
- Spectral-Domain Optical Coherence Tomography with Multiple B-Scan Averaging for Enhanced Imaging of Retinal DiseasesOphthalmology, 2008
- A Review of Results on Sums of Random VariablesActa Applicandae Mathematicae, 2008
- Digital phase stabilization to improve detection sensitivity for optical coherence tomographyMeasurement Science and Technology, 2007
- A new quality assessment parameter for optical coherence tomographyBritish Journal of Ophthalmology, 2006