A comparative study of blood cell count in four automated hematology analyzers: An evaluation of the impact of preanalytical factors
Open Access
- 24 May 2024
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 19 (5) , e0301845
Abstract
Differential white blood cell counts are frequently used in diagnosis, patient stratification, and treatment selection to optimize therapy responses. Referral laboratories are often used but challenged with use of different hematology platforms, variable blood shipping times and storage conditions, and the different sensitivities of specific cell types. To extend the scientific literature and knowledge on the temporal commutability of blood samples between hematology analyzers, we performed a comparative ex-vivo study using four of the most utilized commercial platforms, focusing on the assessment of eosinophils given its importance in asthma management. Whole blood from healthy volunteers with and without atopy (n = 6+6) and participants with eosinophilic asthma (n = 6) were stored under different conditions (at 4, 20, 30, and 37 degrees C, with or without agitation) and analyzed at different time points (3, 6, 24, 48 and 72h post-sampling) in parallel on the Abbott CELL-DYN Sapphire, Beckman Coulter DxH900, Siemens ADVIA 2120i and Sysmex XN-1000V. In the same blood samples, eosinophil-derived neurotoxin (EDN), eosinophil activation and death markers were analyzed. All platforms gave comparable measurements of cell differentials on fresh blood within the same day of sampling. However, by 24 hours, significant temporal and temperature-dependent differences were observed, most markedly for eosinophils. None of the platforms performed perfectly across all temperatures tested during the 72 hours, showing that handling conditions should be optimized depending on the cell type of interest and the hematology analyzer. Neither disease status (healthy vs. asthma) nor agitation of the sample affected the cell quantification result or EDN release. The eosinophil activation markers measured by flow cytometry increased with time, were influenced by temperature, and were higher in those with asthma versus healthy participants. In conclusion, hematology analyzer, time window from sampling until analysis, and temperature conditions must be considered when analyzing blood cell differentials, particularly for eosinophils, via central labs to obtain counts comparable to the values obtained in freshly sampled blood.Keywords
This publication has 14 references indexed in Scilit:
- Performance evaluation of the new hematology analyzer Sysmex XN‐seriesInternational Journal of Laboratory Hematology, 2014
- Spurious white blood cell count from a new automated Sysmex XN hematology analyzerInternational Journal of Laboratory Hematology, 2014
- Comparison of Automated Differential Blood Cell Counts From Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in Normal and Pathologic SamplesAmerican Journal of Clinical Pathology, 2013
- Aging stability of complete blood count and white blood cell differential parameters analyzed by Abbott CELL‐DYN Sapphire hematology analyzerInternational Journal of Laboratory Hematology, 2009
- Stability of hematological analytes depends on the hematology analyser used: A stability study with Bayer Advia 120, Beckman Coulter LH 750 and Sysmex XE 2100Clinica Chimica Acta, 2008
- Automated Blood Cell CountsAmerican Journal of Clinical Pathology, 2008
- Stability of blood cell counts, hematologic parameters and reticulocytes indexes on the Advia A120 hematologic analyzerJournal of Laboratory and Clinical Medicine, 2005
- Quality specification in haematology: the automated blood cell countClinica Chimica Acta, 2004
- The Complete Blood Count: Physiologic Basis and Clinical UsageJournal of Perinatal & Neonatal Nursing, 1997
- Recommendations of the International Council for Standardization in Haematology for Ethylenediaminetetraacetic Acid Anticoagulation of Blood for Blood Cell Counting and Sizing: International Council for Standardization in Haematology: Expert Panel on CytometryAmerican Journal of Clinical Pathology, 1993