Recent Progress on Transcutaneous Energy Transfer for Total Artificial Heart Systems
- 1 May 2001
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 25 (5) , 400-405
Abstract
For many years, transcutaneous energy transfer (TET) systems have been developed for energizing total artificial heart systems. Although such a basic system can be developed without too much design effort, optimization toward high power transfer efficiency forces the introduction of novel system topologies and design strategies. In addition, for medical applications, the thermal impact of a TET system on the biological tissue should be taken into account, resulting in limitations on usable coil geometries. This article presents a TET system that has been developed for a power transfer of 25 W over a distance of 1 cm with minimal dimensions of 1 × 6 × 4 cm for the external driver and 5 × 3 × 1 cm for the internal electronics. The coil geometries have a thickness of 2 mm and a diameter of 6 cm. An overall system efficiency of 80% was achieved for an internal load of 25 W.Keywords
This publication has 5 references indexed in Scilit:
- Recent developments in modeling heat transfer in blood perfused tissuesIEEE Transactions on Biomedical Engineering, 1994
- Use of feedback with voltage regulators for implants powered by coupled coilsMedical & Biological Engineering & Computing, 1985
- System Optimization for Improved Accuracy in Transcutaneous Signal and Power TransmissionIEEE Transactions on Biomedical Engineering, 1984
- Analysis of resonant coupled coils in the design of radio frequency transcutaneous linksMedical & Biological Engineering & Computing, 1983
- Analysis of Tissue and Arterial Blood Temperatures in the Resting Human ForearmJournal of Applied Physiology, 1948