Pulsed inductive microwave magnetometer
Authors:
- 1 October 2002
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 73 (10) , 3563-3569
Abstract
We describe the apparatus, software, and measurement procedures for a pulsed inductive microwave magnetometer (PIMM). PIMM can measure the dynamical properties of materials used in recording heads for magnetic storage applications, and it can be used as a general magnetodynamics diagnostic tool. PIMM uses a coplanar waveguide as both a source of fast pulsed magnetic fields and as an inductive flux sensor. Magnetic field pulses are provided by a 10 V, 55 ps risetime pulse generator; a 20 GHz digital sampling oscilloscope is used to acquire the fast pulse data; and orthogonal Helmholtz pairs provide the bias and saturating fields required for the measurement. The system can measure dynamical behavior as a function of several variables, including applied magnetic bias field, magnetic pulsed field amplitude and width, and sample orientation. Using a fast Fourier transform, PIMM can determine the frequency dependence of the complex magnetic permeability, as well as the step and impulse responses of the magnetic system. Data from 50 nm Ni–Fe and rare-earth-doped Ni–Fe thin films are presented.Keywords
This publication has 9 references indexed in Scilit:
- Control of magnetization dynamics in Ni/sub 81/Fe/sub 19/ thin films through the use of rare-earth dopantsIEEE Transactions on Magnetics, 2001
- High frequency measurements of CoFeHfO thin filmsIEEE Transactions on Magnetics, 2001
- Suppression of magnetic-field pulse-induced magnetization precession by pulse tailoringApplied Physics Letters, 2000
- Frequency- and time-resolved measurements of FeTaN films with longitudinal bias fieldsJournal of Applied Physics, 2000
- Inductive measurement of ultrafast magnetization dynamics in thin-film PermalloyJournal of Applied Physics, 1999
- High speed magnetic recordingIEEE Transactions on Magnetics, 1998
- Direct Observation of Magnetic Relaxation in a Small Permalloy Disk by Time-Resolved Scanning Kerr MicroscopyPhysical Review Letters, 1997
- Free Oscillations of the Magnetization in Permalloy FilmsJournal of Applied Physics, 1961
- Static and Dynamic Behavior of Thin Permalloy FilmsJournal of Applied Physics, 1958