Abstract
Quantitative (q) PCR by amplification of nucleic acid with a fluorescent dye is widely used. Selection of adequate PCR reagents and devices is relevant to achieve reliable and consistent data. Our main objective was to test the robustness of different commercial SYBR green PCR mixes with respect to specificity and sensitivity of the PCR assay, across various PCR machines (Light Cycler 96, ViiA7) and amplification protocols. Herein, we applied PCR protocols for determining mRNA transcript levels, DNA copy numbers, and DNA genotype. First, we set up 70 primer-based assays that targeted immune-related mRNA transcripts. Of the 70 assays 66 (94.3 %) resulted in a single melting curve peak, indicating specificity of the amplification, with PCR mixes from large vendors (Roche, ABI, Bio-Rad). But this was only seen when the PCR protocol that was indicated in the vendor’s guidelines for each particular mix was applied. When deviating from the prescribed protocol, suboptimal melting curves were most often seen when using Roche SYBR green. With respect to PCR yields, the use of ABI mix more often led to lower Cq values. Second, we set up 20 primer-selective PCR assays to target different insertion-deletion and single nucleotide polymorphism regions throughout the genome. The variation in delta Cq between positive and negative DNA samples among the PCR assays was the lowest when using ABI master mix. Finally, the quality of high resolution melting (HRM) assays for DNA genotyping was compared between four commercial HRM PCR mixes (Roche, Bioline, PCR Biosystems, ABI). Only Roche and ABI mixes produced optimal clusters of melting profiles that clearly distinguished genotype variants. The current results show a preference for the use of ABI mix when it comes to obtaining higher sensitivity in cDNA analysis and a higher consistency among assays in distinguishing DNA genotypes among different individuals. For HRM assays, it is advisable to use master mix from a relatively large vendor.Keywords
Funding Information
- Chinese Scholarship Council (201306170038)
This publication has 22 references indexed in Scilit:
- SYBR Green I: Fluorescence Properties and Interaction with DNAJournal of Fluorescence, 2012
- Cross-platform comparison of SYBR® Green real-time PCR with TaqMan PCR, microarrays and other gene expression measurement technologies evaluated in the MicroArray Quality Control (MAQC) studyBMC Genomics, 2008
- Analyzing real-time PCR data by the comparative CT methodNature Protocols, 2008
- Real-Time PCR for mRNA QuantitationBioTechniques, 2005
- Real-time PCR based on SYBR-Green I fluorescence: An alternative to the TaqMan assay for a relative quantification of gene rearrangements, gene amplifications and micro gene deletionsBMC Biotechnology, 2003
- Development of Group- and Serotype-Specific One-Step SYBR Green I-Based Real-Time Reverse Transcription-PCR Assay for Dengue VirusJournal of Clinical Microbiology, 2003
- Gene quantification using real-time quantitative PCRExperimental Hematology, 2002
- Real‐time reverse transcriptase–polymerase chain reaction (RT–PCR) for measurement of cytokine and growth factor mRNA expression with fluorogenic probes or SYBR Green IImmunology & Cell Biology, 2001
- Product Differentiation by Analysis of DNA Melting Curves during the Polymerase Chain ReactionAnalytical Biochemistry, 1997
- Continuous Fluorescence Monitoring of Rapid Cycle DNA AmplificationBioTechniques, 1997