Diagnostic accuracy of the Panbio COVID-19 antigen rapid test device for SARS-CoV-2 detection in Kenya, 2021: A field evaluation
Open Access
- 25 January 2023
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 18 (1) , e0277657
Abstract
Background: Accurate and timely diagnosis is essential in limiting the spread of SARS-CoV-2 infection. The reference standard, rRT-PCR, requires specialized laboratories, costly reagents, and a long turnaround time. Antigen RDTs provide a feasible alternative to rRT-PCR since they are quick, relatively inexpensive, and do not require a laboratory. The WHO requires that Ag RDTs have a sensitivity ≥80% and specificity ≥97%. Methods: This evaluation was conducted at 11 health facilities in Kenya between March and July 2021. We enrolled persons of any age with respiratory symptoms and asymptomatic contacts of confirmed COVID-19 cases. We collected demographic and clinical information and two nasopharyngeal specimens from each participant for Ag RDT testing and rRT-PCR. We calculated the diagnostic performance of the Panbio™ Ag RDT against the US Centers for Disease Control and Prevention’s (CDC) rRT-PCR test. Results: We evaluated the Ag RDT in 2,245 individuals where 551 (24.5%, 95% CI: 22.8–26.3%) tested positive by rRT-PCR. Overall sensitivity of the Ag RDT was 46.6% (95% CI: 42.4–50.9%), specificity 98.5% (95% CI: 97.8–99.0%), PPV 90.8% (95% CI: 86.8–93.9%) and NPV 85.0% (95% CI: 83.4–86.6%). Among symptomatic individuals, sensitivity was 60.6% (95% CI: 54.3–66.7%) and specificity was 98.1% (95% CI: 96.7–99.0%). Among asymptomatic individuals, sensitivity was 34.7% (95% CI 29.3–40.4%) and specificity was 98.7% (95% CI: 97.8–99.3%). In persons with onset of symptoms 7 days (157/876, 17.9%). The highest sensitivity was 87.0% (95% CI: 80.9–91.8%) in symptomatic individuals with cycle threshold (Ct) values ≤30. Conclusion: The overall sensitivity and NPV of the Panbio™ Ag RDT were much lower than expected. The specificity of the Ag RDT was high and satisfactory; therefore, a positive result may not require confirmation by rRT-PCR. The kit may be useful as a rapid screening tool only for symptomatic patients in high-risk settings with limited access to rRT-PCR. A negative result should be interpreted based on clinical and epidemiological information and may require retesting by rRT-PCR.Keywords
Funding Information
- US Centers for Disease Control and Prevention (NU2HGH000032)
This publication has 37 references indexed in Scilit:
- Analytical and clinical performance of the panbio COVID-19 antigen-detecting rapid diagnostic testJournal of Infection, 2021
- Evaluation of the rapid antigen test Panbio COVID-19 in saliva and nasal swabs in a population-based point-of-care studyJournal of Infection, 2021
- Evaluation of a rapid antigen test (Panbio™ COVID-19 Ag rapid test device) for SARS-CoV-2 detection in asymptomatic close contacts of COVID-19 patientsClinical Microbiology and Infection, 2021
- Field evaluation of a rapid antigen test (Panbio™ COVID-19 Ag Rapid Test Device) for COVID-19 diagnosis in primary healthcare centresClinical Microbiology and Infection, 2021
- Panbio antigen rapid test is reliable to diagnose SARS-CoV-2 infection in the first 7 days after the onset of symptomsJournal of Clinical Virology, 2020
- Higher viral loads in asymptomatic COVID-19 patients might be the invisible part of the icebergInfection, 2020
- A cross-country database of COVID-19 testingScientific Data, 2020
- Suppression of a SARS-CoV-2 outbreak in the Italian municipality of Vo’Nature, 2020
- Clearance and persistence of SARS‐CoV‐2 RNA in patients with COVID‐19Journal of Medical Virology, 2020
- The Measurement of Observer Agreement for Categorical DataBiometrics, 1977