Cryogenic multiple reflection absorption cell and Fourier transform spectrometer system for the far infrared
Authors:
- 1 January 1999
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 70 (1) , 23-31
Abstract
A far-infrared system used to measure weak absorption spectra of gases at pressures up to a few atmospheres and at temperatures as low as 20 K is described. The principal component is a multiple reflection mirror cell, cooled either with liquid nitrogen or with the boil-off gas from liquid helium. The cell, of 35 l volume, has an optical beam and a pathlength which can be varied from 4 to 60 m. At the longest pathlength diffraction limits the lowest usable frequency to 20 cm−1. The cell is coupled to a Fourier transform spectrometer and a 1.6 K bolometer. The apparatus has been used in a detailed study of the collision-induced pure translational band of and the translation–rotation bands of and mixtures.
Keywords
This publication has 28 references indexed in Scilit:
- Translational Band of Gaseous Hydrogen at Low TemperatureThe Astrophysical Journal, 1998
- Far-infrared spectrum of N2 and N2-noble gas mixtures near 80 KThe Journal of Chemical Physics, 1996
- The effects of a nonlinear delayed feedback on a chemical reactionThe Journal of Chemical Physics, 1991
- Use of polypropylene film for infrared cryogenic windowsApplied Optics, 1986
- Collision-induced absorption in a gaseous mixture of nitrogen and argonCanadian Journal of Physics, 1986
- The translational–rotational absorption spectrum of hydrogenCanadian Journal of Physics, 1983
- 25-km Low-Temperature Multiple-Reflection CellApplied Optics, 1971
- Laser Pump Cavity with Conical GeometryApplied Optics, 1971
- Far-Infrared Spectra of Ice II, V, and IXThe Journal of Chemical Physics, 1968
- COLLISION-INDUCED ABSORPTION OF COMPRESSED GASES IN THE FAR INFRARED, PART ICanadian Journal of Physics, 1965