Relation of haemofilter type to venous catheter resistance is crucial for filtration performance and haemocompatibility in CVVH—an in vitro study
Authors:
- Juliane K. UngerJUJuliane K. Unger
- Claudia HalternCHClaudia Haltern
- Britta PortzBPBritta Portz
- Bernd DohmenBDBernd Dohmen
- Axel GressnerAGAxel Gressner
- Rolf RossaintRRRolf Rossaint
Open Access
- 27 March 2006
- journal article
- research article
- Published by Oxford University Press (OUP) in Nephrology Dialysis Transplantation
- Vol. 21 (8) , 2191-2201
Abstract
Background. Main factors for the overall performance of haemofilters (HF) are membrane features and filter durability without clogging/clotting of capillaries. However, the venous line resistance (Pv) is a powerful force for net filtrate flux resulting in haemoconcentration and thus is enhancing the phenomenon of membrane clogging. Therefore, we hypothesized that catheter type, as it is associated with Pv-levels, contributes to the extent in which filter longevity and filtration performance are restricted due to blocked hollow fibres. Methods. Heparinized porcine blood (5 IU/ml) was circulated in an in vitro system for haemofiltration (FH6S®—filters were used, Ca. Gambro). Three different sizes of catheters for peripheral vein access (Vygonuele V®, Ca. Vygon) were alternately inserted into the circuit for blood return from the filter to the reservoir. To produce Pv-levels lower than commonly induced by Shaldon catheters, a 14G-vygonuele was used. Pv-levels standard for 11–12 French catheters were provided by using a 16G-vygonuele. To produce Pv-levels common for low-French or tri-lumen catheters, a shortened 18G-vygonuele was used. The respective Pv-levels attained were compared with respect to the overall filtration performance (system pressures, haemocompatibility and sieving coefficients). Results. Catheters of 14 and 16G enabled transiently maximal blood flow (Qb)/filtration rates (Qf) of 300/60 ml/min and continuous filtration with Qb/Qf of 200/40 ml/min. The shortened 18G catheter reduced maximal flow rates down to Qb/Qf of 200/40 ml/min, and continuous flow rates down to Qb/Qf of 125/25 ml/min. At the end, median values for blocked hollow fibres were, 35% in the 14G-group, 40% in the 18G-group and 70% in the 16G-group. Haemocompatibility appeared to be higher in the 14G-group with respect to various parameters when compared with the other groups. Conclusions. The flow resistance by the catheter chosen for haemofiltration clearly influenced the filtration performance. Thus, investigations focused on compatibility of catheter type as it related to Pv-levels with the particular method of renal replacement therapy that should be performed. This point could be crucial in reducing filter clogging and haemostasis during CVVH.Keywords
This publication has 14 references indexed in Scilit:
- HYDROXYETHYL STARCH 130 KD/0.4 AND ALBUMIN IMPROVE CVVH BIOCOMPATIBILITY WHEREAS GELATIN AND HYDROXYETHYL STARCH 200 KD/0.5 LEAD TO ADVERSE SIDE EFFECTS OF CVVH IN ANESTHETIZED PIGSShock, 2006
- Albumin and hydroxyethyl starch 130 kDa/0.4 improve filter clearance and haemocompatibility in haemo- and plasmafiltration—an in vitro studyNephrology Dialysis Transplantation, 2005
- Influence of Different Heparin Concentrations on the Results of In Vitro Investigations in Plasmaseparation Technology Using Capillary Membrane FiltersArtificial Organs, 2003
- Catheter replacement in continuous arteriovenous hemodiafiltration: The balance between infectious and mechanical complications*Critical Care Medicine, 2002
- Prevention of Hemodialysis Catheter-Related Bloodstream Infection Using an Antimicrobial LockBlood Purification, 2002
- AN EX-VIVO EVALUATION OF VASCULAR CATHETERS FOR CONTINUOUS HEMOFILTRATIONRenal Failure, 2002
- Enhancing Filtration Rates by the use of Blood Flow around the Capillaries of Plasmafilters: An in Vitro StudyThe International Journal of Artificial Organs, 2001
- The Influence of Hypoalbuminemia on Maximal Flow Rates and Transmembrane Pressure during Plasmapheresis – An in vitro StudyBlood Purification, 2001
- The coagulation system in the critically ill patient with acute renal failure and the effect of an extracorporeal circuitAmerican Journal of Kidney Diseases, 1997
- A Comparison of the Dual Lumen and Coaxial Catheters for Temporary Hemodialysis AccessThe International Journal of Artificial Organs, 1997