Sonographic Optic Nerve Sheath Diameter as a Screening Tool for Detection of Elevated Intracranial Pressure
Open Access
- 1 January 2013
- journal article
- Vol. 1 (1) , 15-19
Abstract
Timely diagnosis and treatment of post-traumatic elevated intracranial pressure (EICP) could be reduced morbidity and mortality, and improved patients’ outcome. This study is trying to evaluate the diagnostic accuracy of sonographic optic nerve sheath diameter (ONSD) in detection of EICP. Sonographic ONSD of patients with head trauma or cerebrovascular accident suspicious for EICP were evaluated by a trained chief resident of emergency medicine, who was blind to the clinical and brain computed tomography scan (BCT) findings of patients. Immediately after ultrasonography, BCT was performed and reported by an expert radiologist without awareness from other results of the patients. Finally, ultrasonographic and BCT findings regarding EICP were compared. To evaluate the ability of sonographic ONSD in predicting the BCT findings and obtain best cut-off level, receiver operating characteristic (ROC) curve were used. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), positive likelihood ratio (PLR), and negative likelihood ratio (NLR) of sonographic ONSD in determining of EICP was calculated. P < 0.05 was considered as statistically significant. There were 222 patients (65.3% male), with mean age of 42.2±19.5 years (range: 16-90 years). BCT showed signs of EICP, in 28 cases (12.6%). The means of the ONSD in the patients with EICP and normal ICP were 5.5 ± 0.56 and 3.93 ± 0.53 mm, respectively (P<0.0001). ROC curve demonstrated that the best cut off was 4.85 mm. Sensitivity, specificity, PPV, NPV, PLR, and NLR of ONSD for prediction of EICP were 96.4%, 95.3%, 72.2%, 98.9%, 20.6, and 0.04, respectively. Sonographic diameter of optic nerve sheath could be considered as an available, accurate, and noninvasive screening tool in determining the elevated intracranial pressure in cases with head trauma or cerebrovascular accident.Keywords
This publication has 25 references indexed in Scilit:
- Ultrasound measurement of optic nerve sheath diameter in patients with a clinical suspicion of raised intracranial pressureEmergency Medicine Journal, 2010
- Bedside Sonographic Measurement of Optic Nerve Sheath Diameter as a Predictor of Increased Intracranial Pressure in ChildrenAnnals of Emergency Medicine, 2009
- Use of T2-weighted magnetic resonance imaging of the optic nerve sheath to detect raised intracranial pressureCritical Care, 2008
- Optic nerve sonography in the diagnostic evaluation of adult brain injuryCritical Care, 2008
- Correlation of Optic Nerve Sheath Diameter with Direct Measurement of Intracranial PressureAcademic Emergency Medicine, 2008
- Morphometry of the Retrobulbar Human Optic Nerve: Comparison between Conventional Sonography and Ultrafast Magnetic Resonance SequencesInvestigative Ophthalmology & Visual Science, 2007
- The role of optic nerve ultrasonography in the diagnosis of elevated intracranial pressureEmergency Medicine Journal, 2007
- Elevated Intracranial Pressure Detected by Bedside Emergency Ultrasonography of the Optic Nerve SheathAcademic Emergency Medicine, 2003
- Echographic measurement of the optic nerve in patients with intracranial hypertensionmin - Minimally Invasive Neurosurgery, 1987
- Neural and Dural Optic Nerve Measurements With A-Scan UltrasonographySouthern Medical Journal, 1978