Inexpensive Flow Microcalorimeter For Measuring Heat Production of Attached and Sedimentary Aquatic Microorganisms

Authors:

Abstract
A relatively inexpensive flow microcalorimeter is described which is capable of detecting heat outputs as low as 3 .mu.W (precision, .+-. 2%). Its use is illustrated on river epilithon (0.8-6.8 .mu.W cm-2), river sand (9.8 .mu.W cm-3) and marine sand (15.3 .mu.W cm-3); it could be used to detect the heat output from any biotic material over which a flow of water can be passed, provided that such an action would not be disruptive to chemical and biological equilibria.