Proportionality between Wiener spectra of quantum mottle and the squares of modulation transfer functions
- 1 January 1999
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 44 (5) , 1337-1352
Abstract
Rossmann proposed that the Wiener spectra of the quantum mottle of radiographs made using screen-film systems were proportional to the squares of the modulation transfer functions (MTFs) of the screen-film systems. On the other hand, Lubberts theoretically pointed out that the shape of the Wiener spectrum of the quantum mottle depended on the sum of the squares of the MTFs for different depths in the screen phosphor layer, rather than the square of the sum of the MTFs for the different depths, i.e. the square of the MTF of the screen-film systems. The purpose of this study is to experimentally investigate the proportionality between the Wiener spectra of the quantum mottle and the squares of the MTFs of screen-film systems using two screen-film systems having different screen thicknesses. For this purpose, we determined correction factors for the square of the MTF of the screen-film system in the Wiener spectrum of the quantum mottle at each spatial frequency when the Wiener spectral values of the screen mottle were separated into those of the quantum mottle and structure mottle. The correction factor is the ratio of the normalized Wiener spectrum of the quantum mottle to the square of the MTF of the screen-film system. As a result, for a thin screen, the correction factors were unity for all spatial frequencies; on the contrary, for a thick screen, the factor increased with spatial frequency. By calculating the theoretical correction factors using the models for the MTF and Wiener spectrum of the quantum mottle of Nishikawa and Yaffe based on Lubberts' theory, we verified that our experimental results agreed with Lubberts' theory. Furthermore, by obtaining the screen thickness dependence of the theoretical correction factors for the two screens, we showed that, for screens thinner than 0.02 mm, Rossmann's theory can be applied to the relationship between the Wiener spectrum of the quantum mottle and the MTF of the screen-film system, whereas for screens thicker than 0.02 mm, Lubberts' theory should be applied.Keywords
This publication has 20 references indexed in Scilit:
- X‐ray tube voltage dependence of Wiener spectra of quantum mottle obtained by determining detective quantum efficiencies of a screenOptical Engineering, 1996
- X-ray tube voltage dependence of Rossmann's three factors in Wiener spectra of radiographic mottles of a low-speed screen/film system1994
- Fundamental Considerations of Density-Thickness Curves for Soft X-Ray RadiographsThe Journal of Photographic Science, 1993
- Density Dependence of Rossmann’s Three Factors in Wiener Spectra of Radiographic Mottle Made Using Rare-Earth Screen-Film SystemsThe Journal of Photographic Science, 1992
- Model of the spatial‐frequency‐dependent detective quantum efficiency of phosphor screensMedical Physics, 1990
- Effect of finite phosphor thickness on detective quantum efficiencyMedical Physics, 1989
- Radiographic mottle: A comprehensive theoryMedical Physics, 1982
- Image information transfer properties of x‐ray fluorescent screensMedical Physics, 1981
- The effect of phosphor K x-rays on the MTF of rare-earth screensMedical Physics, 1979
- Random Noise Produced by X-Ray Fluorescent Screens*Journal of the Optical Society of America, 1968