Design and evaluation of a miniature laser speckle imaging device to assess gingival health
Open Access
- 1 October 2016
- journal article
- Published by SPIE-Intl Soc Optical Eng in Journal of Biomedical Optics
- Vol. 21 (10) , 1-8
Abstract
Current methods used to assess gingivitis are qualitative and subjective. We hypothesized that gingival perfusion measurements could provide a quantitative metric of disease severity. We constructed a compact laser speckle imaging (LSI) system that could be mounted in custom-made oral molds. Rigid fixation of the LSI system in the oral cavity enabled measurement of blood flow in the gingiva. In vitro validation performed in controlled flow phantoms demonstrated that the compact LSI system had comparable accuracy and linearity compared to a conventional bench-top LSI setup. In vivo validation demonstrated that the compact LSI system was capable of measuring expected blood flow dynamics during a standard postocclusive reactive hyperemia and that the compact LSI system could be used to measure gingival blood flow repeatedly without significant variation in measured blood flow values (p<0.05). Finally, compact LSI system measurements were collected from the interdental papilla of nine subjects and compared to a clinical assessment of gingival bleeding on probing. A statistically significant correlation (ρ=0.53; p<0.005) was found between these variables, indicating that quantitative gingival perfusion measurements performed using our system may aid in the diagnosis and prognosis of periodontal disease.Keywords
Funding Information
- National Institutes of Health (P41 EB015890)
- National Institutes of Health (P41 EB015890)
- National Institute of Health Laser Microbeam Medical Program (P41 EB015890)
- National Institutes of Health Laser Microbeam Medical Program (P41 EB015890)
- National Institutes of Health (R01 DE022831)
- National Institutes of Health (R01 DE022831)
- National Institutes of Health (R01 DE022831)
- Arnold and Mabel Beckman Foundation, the National Institutes of Health (R01 DE022831)
- National Institutes of Health (R01 HD065536)
- National Institutes of Health (R01 HD065536)
- Arnold and Mabel Beckman Foundation, the National Institutes of Health (R01 HD065536)
- National Science Foundation BEST IGERT Program (DGE 1144901)
This publication has 24 references indexed in Scilit:
- Effects of speckle/pixel size ratio on temporal and spatial speckle-contrast analysis of dynamic scattering systems: Implications for measurements of blood-flow dynamicsBiomedical Optics Express, 2013
- Laser speckle contrast imaging for assessment of liver microcirculationMicrovascular Research, 2013
- Longitudinal In Vivo Imaging to Assess Blood Flow and Oxygenation in Implantable Engineered TissuesTissue Engineering Part C: Methods, 2012
- Assessment of pulpal vitality using laser speckle imagingLasers in Surgery and Medicine, 2011
- Laser speckle contrast imaging in biomedical opticsJournal of Biomedical Optics, 2010
- In vivoImaging of Oral Mucositis in an Animal Model Using Optical Coherence Tomography and Optical Doppler TomographyClinical Cancer Research, 2007
- Microvasculature in gingivitis and chronic periodontitis: Disruption of vascular networks with protracted inflammationMicroscopy Research and Technique, 2002
- Dynamic Imaging of Cerebral Blood Flow Using Laser SpeckleJournal of Cerebral Blood Flow & Metabolism, 2001
- Current status of indices of gingivitisJournal of Clinical Periodontology, 1986
- Flow visualization by means of single-exposure speckle photographyOptics Communications, 1981