Single‐mask fabrication of micro‐probe electrode array with various tip heights and sharpness using isotropic and anisotropic etching
Authors:
Open Access
- 1 September 2018
- journal article
- research article
- Published by Institution of Engineering and Technology (IET) in Micro & Nano Letters
- Vol. 13 (9) , 1245-1247
Abstract
In this work, a silicon‐based micro‐probe array was fabricated using multi‐step deep reactive ion etching (DRIE) and reactive ion etching (RIE) processes. The micro‐probe structure was implemented using a micromachining technique through a combination of two anisotropic DRIE and two isotropic RIE processes with a single etch mask. The cone‐shaped tip height and tapered tip‐end angle were experimentally varied with respect to the first DRIE depth. As the first DRIE depth was decreased, the micro‐probe tip height increased and the sharpness consequently improved. The conductive layer was exposed only at the tip ends of the micro‐probe to form a conical ultra‐micro electrode, and it was characterised via cyclic voltammetry measurements.Keywords
This publication has 13 references indexed in Scilit:
- A high-sensitive nano-modified biosensor for dynamic monitoring of glutamate and neural spike covariation from rat cortex to hippocampal sub-regionsJournal of Neuroscience Methods, 2017
- Fabrication of a Micro-Needle Array Electrode by Thermal Drawing for Bio-Signals MonitoringSensors, 2016
- Implantable neurotechnologies: a review of micro- and nanoelectrodes for neural recordingMedical & Biological Engineering & Computing, 2016
- Electrotactile display using microfabricated micro-needle arrayJournal of Micromechanics and Microengineering, 2015
- Conductive microtip electrode array with variable aspect ratio using combination process of reactive ion etchingJournal of Micromechanics and Microengineering, 2013
- A new high-density (25 electrodes/mm2) penetrating microelectrode array for recording and stimulating sub-millimeter neuroanatomical structuresJournal of Neural Engineering, 2013
- Evaluation of the packaging and encapsulation reliability in fully integrated, fully wireless 100 channel Utah Slant Electrode Array (USEA): Implications for long term functionalitySensors and Actuators A: Physical, 2012
- Electrical interfacing between neurons and electronics via vertically integrated sub-4 μm-diameter silicon probe arrays fabricated by vapor–liquid–solid growthBiosensors and Bioelectronics, 2010
- A pseudo 3D glass microprobe array: glass microprobe with embedded silicon for alignment and electrical interconnection during assemblyJournal of Micromechanics and Microengineering, 2009
- An independently addressable microbiosensor array: What are the limits of sensing element density?Faraday Discussions, 2000