Test-Retest Reliability and Preliminary Reference Data for the Virtual Eggs Test in Healthy Adults
Open Access
- 24 February 2026
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Vol. 34 (15344320) , 1363-1370
Abstract
Rehabilitation professionals rely on hand assessment to select the proper therapy and monitor the recovery process over time. The Virtual Eggs Test (VET) is a unique assessment tool that evaluates both fine and gross hand dexterity, taking into account not only the speed in performing an action but also the accuracy in regulating the grip force. The VET was preliminary validated with amputees and healthy individuals, demonstrating construct validity, but also highlighting some weaknesses that required a protocol revision. Here, we present the update version of the protocol and of the outcomes measure (Fine Dexterity Index – FDI - and Gross Dexterity Index - GDI). We carried on a clinical trial involving 79 healthy participants, stratified by age groups, to collect normative data and evaluate test-retest reliability. The FDI and the GDI demonstrate no significant difference between test and retest. FDI showed excellent reliability (ICC was 0.90 and 0.91 for the dominant and non-dominant hand, respectively). While GDI demonstrated moderate reliability (ICC was equal to 0.63 and 0.64 for the dominant and non-dominant hand, respectively). The findings indicate that the VET is reliable over time for the evaluation of hand function in tasks involving fragile objects.Keywords
Funding Information
- Italian Ministry of University and Research through the Complementary Actions to the National Recovery and Resilience Plan (NRRP) “Fit4MedRob—Fit for Medical Robotics” (PNC0000007)
This publication has 48 references indexed in Scilit:
- Improvements on the Virtual Eggs Test to Assess Hand Dexterity2025
- Six‐Month Assessment of a Hand Prosthesis with Intraneural Tactile FeedbackAnnals of Neurology, 2018
- COSMIN guideline for systematic reviews of patient-reported outcome measuresQuality of Life Research, 2018
- Systematic Review of Measures of Impairment and Activity Limitation for Persons With Upper Limb Trauma and AmputationArchives of Physical Medicine and Rehabilitation, 2017
- The COSMIN checklist for assessing the methodological quality of studies on measurement properties of health status measurement instruments: an international Delphi studyQuality of Life Research, 2010
- Manipulation of a fragile objectExperimental Brain Research, 2009
- Repeatability of grip strength and dexterity tests and the effects of age and genderInternational Archives of Occupational and Environmental Health, 2002
- Validation of the Box and Block Test as a measure of dexterity of elderly people: Reliability, validity, and norms studiesArchives of Physical Medicine and Rehabilitation, 1994
- The assessment of hand function: A critical review of techniquesThe Journal of Hand Surgery, 1989
- The assessment and analysis of handedness: The Edinburgh inventoryNeuropsychologia, 1971