Flat detectors and their clinical applications
Author:
- 2 April 2005
- journal article
- physics
- Published by Springer Nature in European Radiology
- Vol. 15 (9) , 1934-1947
Abstract
Diagnostic and interventional flat detector X-ray systems are penetrating the market in all application segments. First introduced in radiography and mammography, they have conquered cardiac and general angiography and are getting increasing attention in fluoroscopy. Two flat detector technologies prevail. The dominating method is based on an indirect X-ray conversion process, using cesium iodide scintillators. It offers considerable advantages in radiography, angiography and fluoroscopy. The other method employs a direct converter such as selenium which is particularly suitable for mammography. Both flat detector technologies are based on amorphous silicon active pixel matrices. Flat detectors facilitate the clinical workflow in radiographic rooms, foster improved image quality and provide the potential to reduce dose. This added value is based on their large dynamic range, their high sensitivity to X-rays and the instant availability of the image. Advanced image processing is instrumental in these improvements and expand the range of conventional diagnostic methods. In angiography and fluoroscopy the transition from image intensifiers to flat detectors is facilitated by ample advantages they offer, such as distortion-free images, excellent coarse contrast, large dynamic range and high X-ray sensitivity. These characteristics and their compatibility with strong magnetic fields are the basis for improved diagnostic methods and innovative interventional applications.Keywords
This publication has 46 references indexed in Scilit:
- Digital detectors for mammography: the technical challengesEuropean Radiology, 2004
- Dose reduction in skeletal and chest radiography using a large-area flat-panel detector based on amorphous silicon and thallium-doped cesium iodide: technical background, basic image quality parameters, and review of the literatureEuropean Radiology, 2004
- Direct-conversion flat-panel X-ray image sensors for digital radiographyProceedings of the IEEE, 2002
- Improved computed radiography image quality from a BaFI:Eu photostimulable phosphor plateMedical Physics, 2002
- The x‐ray sensitivity of amorphous selenium for mammographyMedical Physics, 2002
- Computer-aided diagnosis in chest radiography: a surveyIEEE Transactions on Medical Imaging, 2001
- Digital Radiography of the Skeleton Using a Large-Area Detector Based on Amorphous Silicon Technology: Image Quality and Potential for Dose Reduction in Comparison with Screen-Film RadiographyClinical Radiology, 2000
- In the next decade automated computer analysis will be an accepted sole method to separate “normal” from “abnormal” radiological imagesMedical Physics, 1999
- Flat-Panel X-Ray Detector Using Amorphous Silicon TechnologyInvestigative Radiology, 1997
- Empirical investigation of the signal performance of a high‐resolution, indirect detection, active matrix flat‐panel imager (AMFPI) for fluoroscopic and radiographic operationMedical Physics, 1997