Correlation study between blood pressure and pulse transit time
- 1 February 2015
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-5
Abstract
Blood pressure (BP) determination is a fundamental parameter in cardiovascular assessment. The gold standard method to measure BP is based on the inflatable arm cuff, however has several disadvantages for continuous monitoring. New techniques were developed to overcome these limitations using correlations between the pulse transit time (PTT) and BP. This work draws attention to the PTT rationale using several methods. In order to determine the PTT, an electrocardiogram (ECG) was used combined with multiple photoplethysmography (PPG) sensors applied to different arm locations, these signals were acquired with a bioPLUX device. The Ultrassound system (SonoSite Edge) was used to measure the artery diameter. As reference, BP was measured using a cuff-based sphygmomanometric device. Measurements were performed in a study population of 36 volunteers. The correlation coefficient for DBP determined and DBP measured was r = 0,689. The results suggest PTT deduced from different locations can be used to measure BP.Keywords
This publication has 13 references indexed in Scilit:
- Noninvasive continuous versus intermittent arterial pressure monitoring: evaluation of the vascular unloading technique (CNAP device) in the emergency departmentScandinavian Journal of Trauma, Resuscitation and Emergency Medicine, 2014
- Development of a noninvasive continuous blood pressure measurement and monitoring system2012
- Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based methodEuropean Journal of Applied Physiology, 2011
- Augmentation Index, Left Ventricular Contractility, and Wave ReflectionHypertension, 2009
- Role of Pulse Pressure Amplification in Arterial HypertensionHypertension, 2009
- Effect of confounding factors on blood pressure estimation using pulse arrival timePhysiological Measurement, 2008
- Alterations in human ECG due to the MagnetoHydroDynamic effect: A method for accurate R peak detection in the presence of high MHD artifacts2007
- Indices of cardiovascular function derived from peripheral pulse wave analysis using radial applanation tonometry: a measurement repeatability studyVascular Medicine, 2007
- Continuous non-invasive blood pressure monitoring using concentrically interlocking control loopsComputers in Biology and Medicine, 2006
- Large-artery stiffness, hypertension and cardiovascular risk in older patientsNature Clinical Practice Cardiovascular Medicine, 2005