A rapid field test for the measurement of bovine serum immunoglobulin G using attenuated total reflectance infrared spectroscopy
Open Access
- 20 August 2015
- journal article
- research article
- Published by Springer Nature in BMC Veterinary Research
- Vol. 11 (1) , 218
Abstract
Following the recent development of a new approach to quantitative analysis of IgG concentrations in bovine serum using transmission infrared spectroscopy, the potential to measure IgG levels using technology and a device better designed for field use was investigated. A method using attenuated total reflectance infrared (ATR) spectroscopy in combination with partial least squares (PLS) regression was developed to measure bovine serum IgG concentrations. ATR spectroscopy has a distinct ease-of-use advantage that may open the door to routine point-of-care testing. Serum samples were collected from calves and adult cows, tested by a reference RID method, and ATR spectra acquired. The spectra were linked to the RID-IgG concentrations and then randomly split into two sets: calibration and prediction. The calibration set was used to build a calibration model, while the prediction set was used to assess the predictive performance and accuracy of the final model. The procedure was repeated for various spectral data preprocessing approaches. For the prediction set, the Pearson’s and concordance correlation coefficients between the IgG measured by RID and predicted by ATR spectroscopy were both 0.93. The Bland Altman plot revealed no obvious systematic bias between the two methods. ATR spectroscopy showed a sensitivity for detection of failure of transfer of passive immunity (FTPI) of 88 %, specificity of 100 % and accuracy of 94 % (with IgG <1000 mg/dL as the FTPI cut-off value). ATR spectroscopy in combination with multivariate data analysis shows potential as an alternative approach for rapid quantification of IgG concentrations in bovine serum and the diagnosis of FTPI in calves.Keywords
This publication has 48 references indexed in Scilit:
- Enzyme-Linked Immunosorbent Assay, Single Radial Immunodiffusion, and Indirect Methods for the Detection of Failure of Transfer of Passive Immunity in Dairy CalvesJournal of Veterinary Internal Medicine, 2008
- Use of Fourier-Transform Infrared Spectroscopy for the Diagnosis of Failure of Transfer of Passive Immunity and Measurement of Immunoglobulin Concentrations in HorsesJournal of Veterinary Internal Medicine, 2007
- Hypogammaglobulinemia in Pediatric ICU PatientsJournal of Intensive Care Medicine, 2003
- Passive Transfer of Colostral Immunoglobulins in CalvesJournal of Veterinary Internal Medicine, 2000
- Application of Molecular Spectroscopy in the Mid-Infrared Region to the Determination of Glucose and Cholesterol in Whole Blood and in Blood SerumApplied Spectroscopy, 1997
- Evaluation of 3 Assays for Failure of Passive Transfer in CalvesJournal of Veterinary Internal Medicine, 1996
- Evaluation of a Whole Blood Glutaraldehyde Coagulation Test for the Detection of Failure of Passive Transfer in CalvesJournal of Veterinary Internal Medicine, 1996
- Post-Prandial Blood Glucose Determination by Quantitative Mid-Infrared SpectroscopyApplied Spectroscopy, 1992
- Cross-Validation of Regression ModelsJournal of the American Statistical Association, 1984
- Attenuated total reflectionSpectrochimica Acta, 1961